A viscometric method of measuring plasma fibrinogen concentrations.
Open Access
- 1 June 1992
- journal article
- research article
- Published by BMJ in Journal of Clinical Pathology
- Vol. 45 (6) , 534-535
- https://doi.org/10.1136/jcp.45.6.534
Abstract
A technique based on deducing the viscosity of serum from that of plasma was compared with the commonly used Clauss method. The two methods correlated closely (r = 0.914). The reproducibility of the viscometric method was slightly poorer than the Clauss technique at low fibrinogen concentrations, equal to that at medium fibrinogen concentrations, and marginally better at high concentrations. Fibrinogen can therefore be measured reasonably accurately with the viscometric method, and can be recommended as an alternative for laboratories possessing a viscometer.Keywords
This publication has 6 references indexed in Scilit:
- Plasma fibrinogen — an independent cardiovascular risk factorJournal of Internal Medicine, 1990
- Specific assays of hemostasis proteins: FibrinogenLa Ricerca in Clinica e in Laboratorio, 1990
- PITFALLS IN BLOOD-SAMPLING INFLUENCE OF POSTURE AND VENOUS OCCLUSION TIME1986
- Purification and characterization of 3 fibrinogens with different molecular weights obtained from normal human plasmaThrombosis Research, 1985
- Gerinnungsphysiologische Schnellmethode zur Bestimmung des FibrinogensActa Haematologica, 1957
- THE VISCOSIMETRIC DETERMINATION OF BLOOD FIBRINOGEN1953