Evaluation of metabolic measuring instruments for use in critically ill patients
- 1 June 1990
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 18 (6) , 638-644
- https://doi.org/10.1097/00003246-199006000-00013
Abstract
We evaluted three commerical indirect calorimetry devices which are used during artificial ventilation. Commercial butane, which had an RQ of 0.615, consumes 6.40 ml oxygen, and produces 3.94 ml CO2/1 ml, was burned in a gas-tight combustion chamber in conjunction with the ventilation of a lung model. During combustion, the flow rate of butane was measured with a soapfilm flow meter for the calculation of reference values of oxygen consumption (.ovrhdot.VO2) and CO2 production (.ovrhdot.VCO2). To investigate the effect of oxygen concentration on the accuracy of these instruments, measurements were carried out at FIO2 values of 0.3, 0.4, 0.5, and 0.6 with a fixed ventilation mode (tidal volume 500 ml; respiratory rate 16 breath/min, intermittent positive-pressure ventilation). For the Datex Deltratrac Metabolic Monitor, the mean relative errors of measured .ovrhdot.VO2, .ovrhdot.VCO2, and RQ were all within 4.0%, 2.9%, and 4.0%, respectively. For the Engstrom Metabolic Computer, the corresponding values were 1.4%, 5.7%, and 6.0%, and for the SensorMedics MMC Horizon, 5.7%, 2.9%, and 5.9%.This publication has 10 references indexed in Scilit:
- Validation of oxygen consumption measurements during artificial ventilationJournal of Applied Physiology, 1989
- Measurement of gas exchange in intensive careCritical Care Medicine, 1989
- Evaluation of methods for indirect calorimetry with a ventilated lung modelIntensive Care Medicine, 1989
- Intrapulmonary oxygen consumption in experimental pneumococcal pneumoniaJournal of Applied Physiology, 1988
- Continuous measurement of oxygen uptake Evaluation of the Engström metabolic computer and clinical experiencesActa Anaesthesiologica Scandinavica, 1987
- Evaluation of a microprocessor-controlled exercise testing systemJournal of Applied Physiology, 1984
- Calculation of substrate oxidation rates in vivo from gaseous exchangeJournal of Applied Physiology, 1983
- A Systematic Method for Validation of Gas Exchange MeasurementsAnesthesiology, 1982
- Analysis of error in the determination of respiratory gas exchange at varying FIO2Journal of Applied Physiology, 1981
- An automated system for assessing metabolic and respiratory function during exerciseJournal of Applied Physiology, 1976