Culture with BACTEC Peds Plus/F Bottle Compared with Conventional Methods for Detection of Bacteria in Synovial Fluid
Open Access
- 1 December 2001
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 39 (12) , 4468-4471
- https://doi.org/10.1128/jcm.39.12.4468-4471.2001
Abstract
An evaluation was undertaken to determine the utility of the BACTEC Peds Plus/F bottle and the BACTEC 9240 instrument (Becton Dickinson Diagnostic Instrument Systems, Sparks, Md.) for the detection of clinically significant microorganisms in synovial fluid specimens. The Peds Plus/F bottle was used because in our laboratory the quantity of synovial fluid available for culture is frequently in the range of 0.5 to 3.0 ml. The culture results obtained with the Peds Plus/F bottle were compared to those obtained by a conventional agar plate method for a total of 805 synovial fluid specimens. Microbial growth was produced by 74 cultures (9.2%) from 60 patients, yielding a total of 77 microorganisms. Organisms were classified as pathogens ( n = 62), contaminants ( n = 12), or indeterminate ( n = 3) on the basis of a review of the patients' medical histories. Culture using BACTEC Peds Plus/F bottle detected statistically significantly more pathogens overall (62 versus 51 pathogens [ P = 0.001]) and statistically fewer contaminants overall (1 versus 11 contaminants [ P = 0.006]) than culture by the agar plate method. These results indicate the superior performance of the BACTEC Peds Plus/F bottle over the conventional agar plate method for the detection of clinically significant microorganisms from synovial fluid specimens.Keywords
This publication has 11 references indexed in Scilit:
- Use of the BacT/Alert Blood Culture System for Culture of Sterile Body Fluids Other than BloodJournal of Clinical Microbiology, 1998
- Use of the isolator 1.5 microbial tube for culture of synovial fluid from patients with septic arthritisJournal of Clinical Microbiology, 1997
- Culture of Joint Specimens in Bacterial Arthritis: Impact of blood culture bottle utilizationScandinavian Journal of Rheumatology, 1997
- Comparison of BACTEC Plus 26 and 27 media with and without fastidious organism supplement with conventional methods for culture of sterile body fluidsJournal of Clinical Microbiology, 1994
- High prevalence of Kingella kingae in joint fluid from children with septic arthritis revealed by the BACTEC blood culture systemJournal of Clinical Microbiology, 1992
- Improved method of isolating bacteria from joint fluids by the use of blood culture bottles.Annals of the Rheumatic Diseases, 1986