The Effect of Aortic Valve Incompetence on the Hemodynamics of a Continuous Flow Ventricular Assist Device in a Mock Circulation
- 1 May 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Asaio Journal
- Vol. 54 (3) , 237-244
- https://doi.org/10.1097/mat.0b013e31816a309b
Abstract
There is evidence that the incidence of aortic valve incompetence (AI) and other valvular pathologies may increase as more patients are submitted to longer periods of ventricular assist device (VAD) support. There is a need to better understand the mechanisms associated with the onset of these conditions and other possible complications related to the altered hemodynamics of VAD patients. In this study, the effect of AI on the hemodynamic response of continuous flow VAD (C-VAD) patients was measured in a mock loop over a range of pump speeds and level of native cardiac function. Our results showed that, in the presence of sufficient ventricular function, decreasing the C-VAD speed can allow a transition from series to parallel flow. Our study demonstrated that AI reduces the aortic pressure and flow when system impedance is unchanged. AI produces wasteful recirculation that substantially increases the pump work and decreases systemic perfusion, in particular during series flow conditions coupled with higher C-VAD speeds. The hematologic consequence of this regurgitant flow is a much higher exposure to shear for the blood, increasing the likelihood of hemolysis and thrombosis. While a certain degree of AI can be tolerated by a heart with good cardiac function, the consequences of AI for patients with VADs and poor cardiac function are much greater. Valve dysfunction in VAD patients may be related to structural changes in the tissue induced by altered biomechanics and excessive stress.Keywords
This publication has 28 references indexed in Scilit:
- Quantification of Pulsatility as a Function of Vascular Input Impedance: An In Vitro StudyAsaio Journal, 2007
- Vascular pulsatility in patients with a pulsatile- or continuous-flow ventricular assist deviceThe Journal of Thoracic and Cardiovascular Surgery, 2007
- Effect of Left Ventricular Assist Device Outflow Conduit Anastomosis Location on Flow Patterns in the Native AortaAsaio Journal, 2006
- Native Aortic Valve Insufficiency in Patients With Left Ventricular Assist DevicesThe Annals of Thoracic Surgery, 2006
- Assessment of Recurrent Heart Failure Associated with Left Ventricular Assist Device DysfunctionThe Journal of Heart and Lung Transplantation, 2005
- Cellular and Hemodynamics Responses of Failing Myocardium to Continuous Flow Mechanical Circulatory Support Using the DeBakey-Noon Left Ventricular Assist Device: a Comparative Analysis With Pulsatile-Type DevicesThe Journal of Heart and Lung Transplantation, 2005
- Management of aortic insufficiency in patients with left ventricular assist devices: a simple coaptation stitch method (Park's stitch)The Journal of Thoracic and Cardiovascular Surgery, 2004
- Acquired commissural fusion of aortic valves in patients with left ventricular assist devicesThe Journal of Heart and Lung Transplantation, 2003
- Long-Term Use of a Left Ventricular Assist Device for End-Stage Heart FailureNew England Journal of Medicine, 2001
- Acquired Supravalvular Aortic StenosisThe Annals of Thoracic Surgery, 1979