To report the results of a prospective, nonrandomized, multicenter study of a semiconductor-coated stent in coronary lesions at high risk for stent thrombosis. A balloon-expandable tantalum stent was coated with silicon carbide to enhance thromboresistance (Tensum). Patients were enrolled in an observational study that compared coronary stenting with the Tensum stent in patients at low risk for stent thrombosis against those with factors predisposing to local thrombosis (acute myocardial infarction, small vessel diameter, recanalized chronic total occlusion, saphenous vein bypass grafts, and coronary allograft vascular disease). In 294 patients with 364 coronary lesions, 111 patients with 142 lesions were assigned to the high-risk group. Overall, 406 Tensum stents were implanted (94% procedural success) using antiplatelet medication only after the procedure. The stent thrombosis rate (2.7% overall) in the high-risk group (3.6%) was not significantly different from that of the low-risk patients (2.1%). Silicon carbide coating on coronary stents may inhibit acute/subacute stent thrombosis even in patients at high risk. Randomized trials are underway for further evaluation of this promising coated stent.Keywords
This publication has 9 references indexed in Scilit:
- Intracoronary Stent Implantation Without Ultrasound Guidance and With Replacement of Conventional Anticoagulation by Antiplatelet TherapyCirculation, 1996
- Coronary stent occlusion: Thrombus horribilisJournal of the American College of Cardiology, 1996
- Stents for intracoronary placement: Current status and future directionsJournal of the American College of Cardiology, 1996
- Subacute stent thrombosis: Evolving issues and current conceptsJournal of the American College of Cardiology, 1996
- Heparin-Coated Palmaz-Schatz Stents in Human Coronary ArteriesCirculation, 1996
- Intracoronary Stenting Without Anticoagulation Accomplished With Intravascular Ultrasound GuidanceCirculation, 1995
- A Randomized Comparison of Coronary-Stent Placement and Balloon Angioplasty in the Treatment of Coronary Artery DiseaseNew England Journal of Medicine, 1994
- Physikalische Mechanismen der Festkörper-Protein-Wechselwirkung an der Phasengrenze a-SiC:H-Fibrinogen - Physical Mechanisms of Solid-Protein Interactions at the Interface between Amorphous Silicon Carbide and FibrinogenBiomedizinische Technik/Biomedical Engineering, 1992
- Do stents interfere with the densitometric assessment of a coronary artery lesion?Catheterization and Cardiovascular Diagnosis, 1991