Intraoperative validation of a new system for invasive continuous cardiac output measurement
- 31 January 2009
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 35 (5) , 943-947
- https://doi.org/10.1007/s00134-009-1422-7
Abstract
Although bolus thermodilution technique for cardiac output (CO) measurement has widespread acceptance, new systems are currently available. We evaluated a continuous CO system (TruCCOMS, Aortech International Inc.) that operates on the thermal conservation principle and we compared it with the reference standard transit time flow measurement (TTFM). Nine consecutive cardiac surgery patients were evaluated. After general anesthesia and intubation, a TruCCOMS catheter was percutaneously placed in the pulmonary artery (PA). After median sternotomy and pericardiotomy, a TTFM probe was placed around the main PA. Right ventricular (RV) CO measurements were recorded with both TruCCOMS and TTFM at different times: before cardiopulmonary bypass (CPB) (T0), during weaning from CPB (T1), and prior to sternal closure (T2). Data analysis included paired student t test, Pearson correlation test, and Bland–Altman plotting. TruCCOMS CO values were significantly lower at T0 (TruCCOMS 4.0 ± 1.0 vs. TTFM 4.5 ± 1.0 L/min; P < 0.0001) and T1 (TruCCOMS 3.6 ± 0.5 vs. TTFM 4.2 ± 0.7 L/min; P < 0.0001), and comparable at T2 (TruCCOMS 4.5 ± 0.7 vs. TTFM 4.6 ± 0.8 L/min; P = 0.4). Pearson test showed a significant correlation between TruCCOMS and TTFM CO measurements (RT0 = 0.9, RT1 = 0.8, RT2 = 0.6; P < 0.0001). Bland–Altmann plotting showed a bias of −0.53 ± 0.43 L (−12%) at T0, −0.64 ± 0.43 L (−14.5%) at T1, and −0.1 ± 0.66 L (−0.8%) at T2. Although TruCCOMS may significantly underestimate CO, measurement trends correlate with TTFM. For this reason, a negative trend in RV output should trigger more specific diagnostic procedures.Keywords
This publication has 9 references indexed in Scilit:
- Evaluation of a new continuous cardiac output monitor in off-pump coronary artery surgeryAnaesthesia, 2004
- Original insight into continuous cardiac output monitoring: "TruCCOMS". Correlation with other methods.2003
- Evaluation of a new invasive continuous cardiac output monitoring system: the truCCOMS systemIntensive Care Medicine, 2003
- A Meta-Analysis of Studies Using Bias and Precision Statistics to Compare Cardiac Output Measurement TechniquesJournal of Clinical Monitoring and Computing, 1999
- Analysis of the accuracy of continuous thermodilution cardiac output measurementIntensive Care Medicine, 1996
- Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patientsCritical Care Medicine, 1995
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- Continuous thermodilution cardiac output measurement in intensive care unit patientsJournal of Cardiothoracic and Vascular Anesthesia, 1992