Non‐invasive assessment of cardiac output during exercise in healthy young humans: comparison between Modelflow method and Doppler echocardiography method
- 9 December 2003
- journal article
- research article
- Published by Wiley in Acta Physiologica Scandinavica
- Vol. 179 (4) , 361-366
- https://doi.org/10.1046/j.0001-6772.2003.01211.x
Abstract
Aims: The Modelflow method can estimate cardiac output from arterial blood pressure waveforms using a three‐element model of aortic input impedance (aortic characteristic impedance, arterial compliance, and systemic vascular resistance). We tested the reliability of a non‐invasive cardiac output estimation during submaximal exercise using the Modelflow method from finger arterial pressure waveforms collected by Portapres in healthy young humans. Methods: The Doppler echocardiography method was used as a reference method. Sixteen healthy young subjects (nine males and seven females) performed a multi‐stage cycle ergometer exercise at an intensity corresponding to 70, 90, 110 and 130% of their individual ventilatory threshold for 2 min each. The simultaneous estimation of cardiac output (15 s averaged data) using the Modelflow and Doppler echocardiography methods was performed at rest and during exercise. Results and Conclusion: The Modelflow‐estimated cardiac output correlated significantly with the simultaneous estimates by the Doppler method in all subjects (r = 0.87, P < 0.0001) and the SE of estimation was 1.93 L min−1. Correlation coefficients in each subject ranged from 0.91 to 0.98. Although the Modelflow method overestimated cardiac output, the errors between two estimates were not significantly different among the exercise levels. These results suggest that the Modelflow method using Portapres could provide a reliable estimation of the relative change in cardiac output non‐invasively and continuously during submaximal exercise in healthy young humans, at least in terms of the relative changes in cardiac output.Keywords
This publication has 11 references indexed in Scilit:
- Comparison of Portapres non-invasive blood pressure measurement in the finger with intra-aortic pressure measurement during incremental bicycle exerciseBlood Pressure Monitoring, 2002
- Central and peripheral cardiovascular adaptations to exercise in endurance‐trained childrenActa Physiologica Scandinavica, 2002
- Non‐invasive cardiac output assessment during moderate exercise: pulse contour compared with CO2 rebreathingClinical Physiology and Functional Imaging, 1999
- Test–Retest Reproducibility of Maximum Cardiac Output by Doppler EchocardiographyThe American Journal of Cardiology, 1998
- Effects of endurance training on the size and blood flow of the arterial conductance vessels in humansActa Physiologica Scandinavica, 1998
- Noninvasive cardiac output measurement by arterial pulse analysis compared with inert gas rebreathingJournal of Applied Physiology, 1993
- Continuous noninvasive cardiac output as estimated from the pulse contour curveJournal of Clinical Monitoring and Computing, 1992
- Determination of stroke volume and cardiac output during exercise: comparison of two-dimensional and Doppler echocardiography, Fick oximetry, and thermodilution.Circulation, 1987
- A new method for detecting anaerobic threshold by gas exchangeJournal of Applied Physiology, 1986
- The static elastic properties of 45 human thoracic and 20 abdominal aortas in vitro and the parameters of a new modelJournal of Biomechanics, 1984