Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations
Top Cited Papers
- 24 October 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (22) , 5785-5807
- https://doi.org/10.1088/0031-9155/51/22/005
Abstract
A study of the performance of five commercial radiotherapy treatment planning systems (TPSs) for common treatment sites regarding their ability to model heterogeneities and scattered photons has been performed. The comparison was based on CT information for prostate, head and neck, breast and lung cancer cases. The TPSs were installed locally at different institutions and commissioned for clinical use based on local procedures. For the evaluation, beam qualities as identical as possible were used: low energy (6 MV) and high energy (15 or 18 MV) x-rays. All relevant anatomical structures were outlined and simple treatment plans were set up. Images, structures and plans were exported, anonymized and distributed to the participating institutions using the DICOM protocol. The plans were then re-calculated locally and exported back for evaluation. The TPSs cover dose calculation techniques from correction-based equivalent path length algorithms to model-based algorithms. These were divided into two groups based on how changes in electron transport are accounted for ((a) not considered and (b) considered). Increasing the complexity from the relatively homogeneous pelvic region to the very inhomogeneous lung region resulted in less accurate dose distributions. Improvements in the calculated dose have been shown when models consider volume scatter and changes in electron transport, especially when the extension of the irradiated volume was limited and when low densities were present in or adjacent to the fields. A Monte Carlo calculated algorithm input data set and a benchmark set for a virtual linear accelerator have been produced which have facilitated the analysis and interpretation of the results. The more sophisticated models in the type b group exhibit changes in both absorbed dose and its distribution which are congruent with the simulations performed by Monte Carlo-based virtual accelerator.Keywords
This publication has 49 references indexed in Scilit:
- Large efficiency improvements in BEAMnrc using directional bremsstrahlung splittingMedical Physics, 2004
- A virtual linear accelerator for verification of treatment planning systemsPhysics in Medicine & Biology, 2000
- Calculation models for determining the absorbed dose in water phantoms in off-axis planes of rectangular fields of open and wedged photon beamsPhysics in Medicine & Biology, 1995
- Investigation of the convolution method for polyenergetic spectraMedical Physics, 1993
- Acquisition of the effective lateral energy fluence distribution for photon beam dose calculations by convolution modelsPhysics in Medicine & Biology, 1991
- Generation of photon energy deposition kernels using the EGS Monte Carlo codePhysics in Medicine & Biology, 1988
- Verification by Monte Carlo methods of a power law tissue-air ratio algorithm for inhomogeneity corrections in photon beam dose calculationsPhysics in Medicine & Biology, 1980
- Clinical application of a CT based treatment planning systemComputerized Tomography, 1978
- Program IRREG — Calculation of dose from irregularly shaped radiation beamsComputer Programs in Biomedicine, 1972
- Scatter-air ratiosPhysics in Medicine & Biology, 1972