A virtual linear accelerator for verification of treatment planning systems
- 21 September 2000
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (10) , 2887-2896
- https://doi.org/10.1088/0031-9155/45/10/311
Abstract
A virtual linear accelerator is implemented into a commercial pencil-beam-based treatment planning system (TPS) with the purpose of investigating the possibility of verifying the system using a Monte Carlo method. The characterization set for the TPS includes depth doses, profiles and output factors, which is generated by Monte Carlo simulations. The advantage of this method over conventional measurements is that variations in accelerator output are eliminated and more complicated geometries can be used to study the performance of a TPS. The difference between Monte Carlo simulated and TPS calculated profiles and depth doses in the characterization geometry is less than ±2% except for the build-up region. This is of the same order as previously reported results based on measurements. In an inhomogeneous, mediastinum-like case, the deviations between TPS and simulations are small in the unit-density regions. In low-density regions, the TPS overestimates the dose, and the overestimation increases with increasing energy from 3.5% for 6 MV to 9.5% for 18 MV. This result points out the widely known fact that the pencil beam concept does not handle changes in lateral electron transport, nor changes in scatter due to lateral inhomogeneities. It is concluded that verification of a pencil-beam-based TPS with a Monte Carlo based virtual accelerator is possible, which facilitates the verification procedure.Keywords
This publication has 21 references indexed in Scilit:
- Verification of a pencil beam based treatment planning system: output factors for open photon beams shaped with MLC or blocksPhysics in Medicine & Biology, 1999
- Measured lung dose correction factors for 50 MV photonsPhysics in Medicine & Biology, 1998
- American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planningMedical Physics, 1998
- Evaluation of optimized compensators on a 3D planning systemMedical Physics, 1998
- Limitations of a pencil beam approach to photon dose calculations in the head and neck regionRadiotherapy and Oncology, 1995
- A pencil beam model for photon dose calculationMedical Physics, 1992
- Monte Carlo techniques in medical radiation physicsPhysics in Medicine & Biology, 1991
- Acquisition of the effective lateral energy fluence distribution for photon beam dose calculations by convolution modelsPhysics in Medicine & Biology, 1991
- Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous mediaMedical Physics, 1989
- Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986