In vitro comparison of the two hard-hard articulations for total hip replacements
Top Cited Papers
- 1 February 2001
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 215 (2) , 153-160
- https://doi.org/10.1243/0954411011533715
Abstract
Polyethylene particle disease is one of the major causes of late aseptic loosening of total hip replacement. Two hard-hard articulations (alumina-on-alumina and metal-on-metal) have been developed in Europe as an alternative to the ultra-high molecular weight polyethylene (UHMWPE) articulations. Even though these hard-hard articulations are on the market and numerous reports have been published about them, only a very limited number of studies allowing a direct in vitro comparison of the two articulations have been published so far. This paper compares in vitro these two types of articulation (alumina-on-alumina and metal-on-metal), which have been tested with a hip simulator for their tribological behaviour using exactly the same experimental methodology. This comparison shows that these two types of hard-hard articulation have very similar abrasive wear behaviour with four main features: 1. A running-in wear period (1 × 106 cycles) gives a cumulative wear of about 20 μm with head diameters of 28 mm. 2. After the running-in wear, there is a stabilization of the linear wear behaviour with a low linear wear rate/106 cycles for both types of articulation. 3. The volumetric wear rate of both articulations (3/year for head diameters of 28mm) is significantly lower than that observed for metal-on-polyethylene or ceramic-on-polyethylene articulations having the same head diameter. 4. Abrasive wear is readily apparent (indicating a mixed lubrication regime) with both types of articulation. The extremely low wear performance of these articulations is confirmed and they constitute a lowwear alternative to the UHMWPE articulations currently used.Keywords
This publication has 14 references indexed in Scilit:
- Migration, Particles, and Fluid PressurePublished by Wolters Kluwer Health ,1998
- Low wear and friction in alumina/alumina total hip joints: A hip simulator studyActa Orthopaedica, 1998
- Clinical Wear Performance of Metal-on-Metal Hip ArthroplastiesPublished by ASTM International ,1998
- Simulator Wear of Some Commercially Available Metal-on-Metal Hip ImplantsPublished by ASTM International ,1998
- The Role of Access of Joint Fluid to Bone in Periarticular OsteolysisJournal of Bone and Joint Surgery, 1997
- Wear Behavior and Histopathology of Classic Cemented Metal on Metal Hip EndoprosthesesClinical Orthopaedics and Related Research, 1996
- Metal on Metal Bearings in Hip ArthroplastyClinical Orthopaedics and Related Research, 1996
- Wear analysis of retrieved alumina heads and sockets of hip prosthesesJournal of Biomedical Materials Research, 1989
- BiorheologieThe Science of Nature, 1983
- The Walking Activity of Patients with Artificial Hip JointsEngineering in Medicine, 1982