Factors influencing flip angle mapping in MRI: RF pulse shape, slice‐select gradients, off‐resonance excitation, and B0 inhomogeneities
- 13 June 2006
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 56 (2) , 463-468
- https://doi.org/10.1002/mrm.20947
Abstract
To understand the various effects that influence actual flip angles, and correct for these effects, it is important to precisely quantify the MRI parameters (such as T1, T2, and perfusion). In this paper actual flip angle maps are calculated using a conventional gradient-echo (GRE) sequence with different radiofrequency (RF) pulse shapes (Gaussian, sinc, and truncated-sinc), slice-selection gradients, off-resonance excitations, and B0 field inhomogeneities. The experimental results demonstrate that RF pulse shapes significantly affect the flip angle distribution and calibration factors. Off-resonance RF excitations, B0 nonuniformities, and slice-selection gradients can lead to degradations in the signal intensities of the images used to map the flip angle, and potentially introduce a bias and increased variance in the measured flip angles.Keywords
This publication has 15 references indexed in Scilit:
- In vivo method for correcting transmit/receive nonuniformities with phased array coilsMagnetic Resonance in Medicine, 2005
- Measurement and correction of transmitter and receiver induced nonuniformities in vivoMagnetic Resonance in Medicine, 2005
- Fast robust automated brain extractionHuman Brain Mapping, 2002
- Adiabatic radiofrequency pulse forms in biomedical nuclear magnetic resonanceConcepts in Magnetic Resonance, 2002
- Imaging of the active B1 field in vivoMagnetic Resonance in Medicine, 1996
- High frequency volume coils for clinical NMR imaging and spectroscopyMagnetic Resonance in Medicine, 1994
- Effects of RF amplifier distortion on selective excitation and their correction by prewarpingMagnetic Resonance in Medicine, 1992
- Variable excitation angle AFP pulsesMagnetic Resonance in Medicine, 1989
- Amplitude- and frequency/phase-modulated refocusing pulses that induce plane rotations even in the presence of inhomogeneous B1 fieldsJournal of Magnetic Resonance (1969), 1988
- The intrinsic signal‐to‐noise ratio in NMR imagingMagnetic Resonance in Medicine, 1986