Artefact reduction in dual-radionuclide subtraction studies
- 1 April 1987
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 32 (4) , 477-493
- https://doi.org/10.1088/0031-9155/32/4/006
Abstract
A method is proposed which significantly reduces the artefacts commonly experienced in dual radionuclide subtraction studies. Images of two radionuclides recorded simultaneously differ in resolution, sensitivity and attenuation. Also, one image will include scatter from the second higher-energy radionuclide. As a result severe artefacts are likely to occur when the two images are subtracted. In order to minimise the depth dependence of resolution, attenuation and scatter, the geometric mean of conjugate views was considered. From experimental work with activity placed in a depth of water it was demonstrated that the number and spatial distribution of scattered photons recorded in any energy window could be accurately predicted from the geometric mean image recorded in the photopeak. This prediction was accurate, independent of the depth of the source in water for a range of phantom dimensions. Differences in the instrument sensitivity and resolution at different energies can also be readily compensated for by using geometric mean images, as can differences due to the variation in attenuation. In practice three factors can be experimentally determined for any pair of radionuclides: a scatter ratio, a scatter function and a resolution compensation function. These data are then used to improve the dual-radionuclide subtraction analysis. The ability of the technique to significantly reduce subtraction artefacts has been demonstrated in phantom studies.This publication has 10 references indexed in Scilit:
- Parathyroid adenomas evaluated by Tl-201/Tc-99m pertechnetate subtraction scintigraphy and high-resolution ultrasonography.Radiology, 1985
- Scatter Compensation Techniques for SPECTIEEE Transactions on Nuclear Science, 1985
- Dual isotope scintigraphy in stroke patientsNeuroradiology, 1984
- The limitations of the dual radionuclide subtraction technique for the external detection of tumours by radioiodine-labelled antibodiesThe British Journal of Radiology, 1983
- NON-INVASIVE DETECTION OF CORONARY THROMBI WITH IN-111 PLATELETS - CONCISE COMMUNICATION1983
- The measurement of the depth of a point source of a radioisotope from gamma ray spectraPhysics in Medicine & Biology, 1979
- A technique for the absolute measurement of activity using a gamma camera and computerPhysics in Medicine & Biology, 1979
- Performance characteristics of a gamma camera over a wide range of energiesEuropean Journal of Nuclear Medicine and Molecular Imaging, 1978
- In vivo quantitation of lesion radioactivity using external counting methodsMedical Physics, 1976
- Regional Cerebral Blood Flow Estimated by 133 Xenon InhalationStroke, 1975