Stent Appearance at Contrast-enhanced MR Angiography: In Vitro Examination with 14 Stents
- 1 October 2000
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 217 (1) , 173-178
- https://doi.org/10.1148/radiology.217.1.r00se28173
Abstract
To evaluate signal intensity changes influencing assessment of stent patency at contrast material-enhanced magnetic resonance (MR) angiography. By using an in vitro model, 14 stents-nine nitinol, one tantalum, two stainless steel, and two cobalt alloy-were investigated regarding their appearance at MR imaging. A vascular phantom consisting of tubes filled with 2.00 mmol/L gadopentetate dimeglumine in saline solution was studied in different orientations within the magnetic field. Imaging was performed with a fast three-dimensional gradient-echo sequence (4. 70/1.89 [repetition time msec/echo time msec]). Relative signal intensity reduction within the stents and the degree of artificial narrowing of the stent lumen were calculated. The stent lumen was visible within 13 stents. A total signal void inside the stent lumen appeared in only one cobalt alloy stent. Artificial narrowing of the diameter was less than 33% in 10 of 14 stents. The tantalum stent and four nitinol stents seemed best suited for contrast-enhanced MR angiography. A bandlike artifact occurred at the ends of the stents when positioned along the readout direction. To differentiate between artifacts and stenoses, knowledge of the degree of signal intensity reduction and artificial lumen narrowing within vascular stents is essential. Stent geometry, relative orientation to the magnetic field, and alloy composition influence signal intensity alteration within the stent lumen.Keywords
This publication has 20 references indexed in Scilit:
- Accessory equipment considerations with respect to MRI compatibilityJournal of Magnetic Resonance Imaging, 1998
- Interventional MRI at high‐field (1.5 T): Needle artifactsJournal of Magnetic Resonance Imaging, 1998
- Evaluation of abdominal aortic aneurysm for stent-graft placement: comparison of gadolinium-enhanced MR angiography versus helical CT angiography and digital subtraction angiography.Radiology, 1997
- Artefakte durch Dentallegierungen in der MR-Bildgebung des Temporomandibulargelenks: eine Phantommodellstudie bei 1,5 TRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1997
- Visualization of magnetic resonance-compatible needles at 1.5 and 0.2 TeslaCardioVascular and Interventional Radiology, 1996
- Breath-hold gadolinium-enhanced MR angiography of the abdominal aorta and its major branches.Radiology, 1995
- Coronary MR angiography after coronary stent placement.American Journal of Roentgenology, 1995
- Gadolinium Enhanced MR Imaging of Vascular StentsJournal of Computer Assisted Tomography, 1990
- Tantalum vascular stents: in vivo evaluation with MR imaging.Radiology, 1989
- MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils.Radiology, 1988