Electrodynamics and ultimate SNR in parallel MR imaging
Top Cited Papers
Open Access
- 23 July 2004
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 52 (2) , 376-390
- https://doi.org/10.1002/mrm.20183
Abstract
The purpose of this article is to elucidate inherent limitations to the performance of parallel MRI. The study focuses on the ultimate signal-to-noise ratio (SNR), which refers to the maximum SNR permitted by the electrodynamics of the signal detection process. Using a spherical model object, it is shown that the behavior of the ultimate SNR imposes distinct limits on the acceleration rate in parallel imaging. For low and moderate acceleration, the ultimate SNR performance is nearly optimal, with geometry factors close to 1. However, for high reduction factors beyond a critical value, the ultimate performance deteriorates rapidly, corresponding to exponential growth of the geometry factor. The transition from optimal to deteriorating performance depends on the electrodynamic characteristics of the detected RF fields. In the near-field regime, i.e., for low B0 and small object size, the critical reduction factor is constant and approximately equal to four for 1D acceleration in the sphere. In the far-field wave regime the critical reduction factor is larger and increases both with B0 and object size. Therefore, it is concluded that parallel techniques hold particular promise for human MR imaging at very high field. Magn Reson Med 52:376–390, 2004.Keywords
This publication has 22 references indexed in Scilit:
- Ultimate intrinsic signal‐to‐noise ratio for parallel MRI: Electromagnetic field considerationsMagnetic Resonance in Medicine, 2003
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Design of a SENSE‐optimized high‐sensitivity MRI receive coil for brain imagingMagnetic Resonance in Medicine, 2002
- Specific coil design for SENSE: A six-element cardiac arrayMagnetic Resonance in Medicine, 2001
- Sensitivity profiles from an array of coils for encoding and reconstruction in parallel (SPACE RIP)Magnetic Resonance in Medicine, 2000
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- AUTO-SMASH: A self-calibrating technique for SMASH imagingMagnetic Resonance Materials in Physics, Biology and Medicine, 1998
- Ultimate intrinsic signal‐to‐noise ratio in MRIMagnetic Resonance in Medicine, 1998
- Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arraysMagnetic Resonance in Medicine, 1997
- The signal-to-noise ratio of the nuclear magnetic resonance experimentJournal of Magnetic Resonance (1969), 1976