Signal‐to‐noise ratio and spectral linewidth improvements between 1.5 and 7 Tesla in proton echo‐planar spectroscopic imaging
Open Access
- 8 November 2006
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 56 (6) , 1200-1210
- https://doi.org/10.1002/mrm.21067
Abstract
This study characterizes gains in sensitivity and spectral resolution of proton echo‐planar spectroscopic imaging (PEPSI) with increasing magnetic field strength (B0). Signal‐to‐noise ratio (SNR) per unit volume and unit time, and intrinsic linewidth (LW) of N‐acetyl‐aspartate (NAA), creatine (Cr), and choline (Cho) were measured with PEPSI at 1.5, 3, 4, and 7 Tesla on scanners that shared a similar software and hardware platform, using circularly polarized (CP) and eight‐channel phased‐array (PA) head coils. Data were corrected for relaxation effects and processed with a time‐domain matched filter (MF) adapted to each B0. The SNR and LW measured with PEPSI were very similar to those measured with conventional point‐resolved spectroscopy (PRESS) SI. Measurements with the CP coil demonstrated a nearly linear SNR gain with respect to B0 in central brain regions. For the PA coil, the SNR–B0 relationship was less than linear, but there was a substantial SNR increase in comparison to the CP coil. The LW in units of ppm decreased with B0, resulting in improved spectral resolution. These studies using PEPSI demonstrated linear gains in SNR with respect to B0, consistent with theoretical expectations, and a decrease in ppm LW with increasing B0. Magn Reson Med, 2006.Keywords
This publication has 35 references indexed in Scilit:
- Time-domain semi-parametric estimation based on a metabolite basis setNMR in Biomedicine, 2005
- Parallel imaging performance as a function of field strength—An experimental investigation using electrodynamic scalingMagnetic Resonance in Medicine, 2004
- Comparison of longitudinal metabolite relaxation times in different regions of the human brain at 1.5 and 3 TeslaMagnetic Resonance in Medicine, 2003
- Proton echo-planar spectroscopic imaging with highly effective outer volume suppression using combined presaturation and spatially selective echo dephasingMagnetic Resonance in Medicine, 2003
- Proton T1 and T2 relaxation times of human brain metabolites at 3 TeslaNMR in Biomedicine, 2001
- Proton NMR chemical shifts and coupling constants for brain metabolitesNMR in Biomedicine, 2000
- Noise in MRIMagnetic Resonance in Medicine, 1996
- Analytical resolution and noise characteristics of linearly reconstructed magnetic resonance data with arbitraryk‐space samplingMagnetic Resonance in Medicine, 1995
- The intrinsic signal‐to‐noise ratio in NMR imagingMagnetic Resonance in Medicine, 1986
- The sensitivity of the zeugmatographic experiment involving human samplesJournal of Magnetic Resonance (1969), 1979