In vitro simulation of contact fatigue damage found in ultra-high molecular weight polyethylene components of knee prostheses
- 1 April 1998
- journal article
- research article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 212 (4) , 293-302
- https://doi.org/10.1243/0954411981534079
Abstract
An in vitro simulation of fatigue loading of ultra-high molecular weight polyethylene (UHMWPE) knee components was carried out on a knee simulator and on a rolling and sliding wear tester. Tibial components for the knee simulator were gamma-sterilized, implantable components taken from manufacturing inventory. The rolling/sliding UHMWPE discs were machined from bar stock and either gamma sterilized in air and accelerated aged, or left as non-sterilized (controls). Cracking and delamination of samples that had been gamma sterilized in air and aged were observed in both types of tests. Contact fatigue damage was visible in as few as 150 000 cycles using the knee simulator at loads of 122 N (275 lb). The rolling/sliding samples showed signs of damage in as few as 130 000 cycles with an estimated stress of 15 MPa and 25 per cent sliding. However, cracking and delamination were not generated in the never-sterilized or recently sterilized controls. UHMWPE that has been gamma sterilized in air and aged is shown to be susceptible to contact fatigue damage. These results are important to the interpretation of in vitro total knee replacement simulations used to assess the performance of tibial bearings.Keywords
This publication has 22 references indexed in Scilit:
- Wear testing of materials and surfaces for total knee replacementJournal of Biomedical Materials Research, 1996
- Finite Element Analysis of a Layered Elastic Solid in Normal Contact With a Rigid SurfaceJournal of Tribology, 1988
- The Problem of Surface Damage in Polyethylene Total Knee ComponentsClinical Orthopaedics and Related Research, 1986
- Fatigue crack propagation behavior of ultrahigh molecular weight polyethyleneJournal of Orthopaedic Research, 1984
- Retrieval analysis of total knee prostheses: A method and its application to 48 total condylar prosthesesJournal of Biomedical Materials Research, 1983
- Wear of Artificial Joint Materials IIIEngineering in Medicine, 1981
- The Deformation and Wear of Plastic Components in Artificial Knee Joints — An Experimental StudyEngineering in Medicine, 1981
- The appearances of wear on polyethylene—a comparison of in vivo and in vitro wear surfacesJournal of Biomedical Materials Research, 1978
- Elasto-Plastic Indentation of a Layered MediumJournal of Engineering Materials and Technology, 1974