A Monte Carlo investigation of fluence profiles collimated by an electron specific MLC during beam delivery for modulated electron radiation therapy
- 17 October 2002
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 29 (11) , 2472-2483
- https://doi.org/10.1118/1.1513160
Abstract
Modulated electron radiation therapy (MERT) is able to deliver conformal dose to shallow tumors while significantly reducing dose to distal structures and surrounding tissues. An electron specific multileaf collimator (eMLC) has been proposed and constructed as an effective means of delivering electron beams for MERT. The aim of this work is to apply the Monte Carlo method to investigate the fluence profiles collimated by the eMLC in order to achieve accurate beam delivery for MERT. In this work, theEGS4/BEAMcode was used to simulate the eMLC collimated electron beams of 6–20 MeV generated from a Varian Clinac 2100C linear accelerator. An attempt was made to describe the fluence profiles with an analytic Sigmoid function. The function parameters were determined by the fittings of the Monte Carlo simulated fluence profiles. How the function parameters depend on the eMLC aperture size, the off‐axis location, and the electron beam energy has been investigated. It has been found that the eMLC collimated fluence profiles are dependent on beam energy, while almost independent of leaf location or dimension of MLC aperture. There is little difference in the fluence profiles collimated by the leaf side and the leaf end for the straight‐edged leaves. It is possible that these energy‐dependent Sigmoid functions can serve as operators to account for the energy dependence of the eMLC collimated fluence profiles. These operators can be incorporated into the inverse planning algorithm to derive desired dose distributions using a set of electron beams of variable energy and field size suitable for delivery by the eMLC.Keywords
This publication has 20 references indexed in Scilit:
- Modulated electron radiation therapy: A new treatment modalityInternational Journal of Radiation Oncology*Biology*Physics, 2000
- Monte Carlo and experimental investigations of multileaf collimated electron beams for modulated electron radiation therapyMedical Physics, 2000
- Treatment head design for multileaf collimated high-energy electronsMedical Physics, 1999
- Feasibility study of multileaf collimated electrons with a scattering foil based acceleratorRadiotherapy and Oncology, 1996
- Matching of electron beams for conformal therapy of target volumes at moderate depthsRadiotherapy and Oncology, 1996
- Electron dose profile shaping by modulation of a scanning elementary beamMedical Physics, 1996
- A model for electron‐beam applicator scatterMedical Physics, 1995
- A Monte Carlo investigation of electron-beam applicator scatterMedical Physics, 1995
- Design Principles and Clinical Possibilities with A New Generation of Radiation Therapy Equipment: A ReviewActa Oncologica, 1987
- Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986