Urinary oxalate determination by use of immobilized oxalate oxidase in a continuous-flow system.
Open Access
- 1 January 1983
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 29 (1) , 16-20
- https://doi.org/10.1093/clinchem/29.1.16
Abstract
In this procedure, oxalate oxidase (EC 1.2.3.4) immobilized in a continuous-flow system is used to determine oxalate in urine. The hydrogen peroxide formed from oxalate is detected by use of a color reaction with peroxidase (EC 1.11.1.7), 3-methyl-2-benzothiazoline hydrazine, and N,N-dimethylalanine. However, urine contains an oxalate oxidase inhibitor, which cannot be removed by heating, ion-exchange resins, or cellulose columns. This makes it necessary to precipitate the oxalate before assay. The overall assay system is accurate (oxalate recovery, 95.9%), sensitive (less than or equal to 5 mumol/L), precise (within-batch CV less than 1.25%, between-batch CV less than 5%), and relatively rapid (60 samples per working day). The assay system has better accuracy than an established chemical method and a gas-chromatographic method, and is considerably less arduous than and correlates well (r = 0.94) with a modified chemical method. The reference interval for urinary oxalate excretion is 0.16-0.56 mmol per day (n = 97). Only nonphysiological concentrations of ascorbate interfere with the assay, by increasing the oxalate result in the overall assay, presumably by post-micturition formation of oxalate from ascorbate in the urine samples.This publication has 8 references indexed in Scilit:
- The relationship of the urinary ascorbate metabolites to specific levels of ascorbate supplementation in the monkeyThe American Journal of Clinical Nutrition, 1981
- Direct Spectrophotometric Determination of Serum and Urinary Oxalate with Oxalate Oxidasecclm, 1981
- THE PRODUCTION OF |14C| OXALATE DURING THE METABOLISM OF |14C| CARBOHYDRATES IN ISOLATED RAT HEPATOCYTESImmunology & Cell Biology, 1980
- SPECTROPHOTOMETRIC DETERMINATION OF URINARY OXALATE WITH OXALATE OXIDASE PREPARED FROM MOSS1980
- Gas-chromatographic estimation of urinary oxalate and its comparison with a colorimetric method.Clinical Chemistry, 1979
- Immobilized glucose oxidase used in the continuous-flow determination of serum glucose.Clinical Chemistry, 1976
- Purification and properties of oxalic acid oxidaseArchives of Biochemistry and Biophysics, 1966
- A study of the inhibition of catalase by 3-amino-1:2:4-triazoleBiochemical Journal, 1958