Analysis of the spatial and temporal accuracy of heating in the prostate gland using transurethral ultrasound therapy and active MR temperature feedback
- 8 April 2009
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (9) , 2615-2633
- https://doi.org/10.1088/0031-9155/54/9/002
Abstract
A new MRI-guided therapy is being developed as a minimally invasive treatment for localized prostate cancer utilizing high-intensity ultrasound energy to generate a precise region of thermal coagulation within the prostate gland. The purpose of this study was to evaluate in vivo the capability to produce a spatial heating pattern in the prostate that accurately matched the shape of a target region using transurethral ultrasound heating and active MR temperature feedback. Experiments were performed in a canine model (n = 9) in a 1.5 TMR imager using a prototype device comprising a single planar transducer operated under rotational control. The spatial temperature distribution, measured every 5 s with MR thermometry, was used to adjust the acoustic power and rotation rate in order to achieve a temperature of 55 degrees C along the outer boundary of the target region. The results demonstrated the capability to produce accurate spatial heating patterns within the prostate gland. An average temperature of 56.2 +/- 0.6 degrees C was measured along the outer boundary of the target region across all experiments in this study. The average spatial error between the target boundary and the 55 degrees C isotherm was 0.8 +/- 0.7 mm (-0.2 to 3.2 mm), and the overall treatment time was <= 20 min for all experiments. Excellent spatial agreement was observed between the temperature information acquired with MRI and the pattern of thermal damage measured on H&E-stained tissue sections. This study demonstrates the benefit of adaptive energy delivery using active MR temperature feedback, and an excellent capability to treat precise regions within the prostate gland with this technology.Keywords
This publication has 43 references indexed in Scilit:
- MRI‐compatible transurethral ultrasound system for the treatment of localized prostate cancer using rotational controlMedical Physics, 2008
- Prostate Tissue Analysis Immediately Following Magnetic Resonance Imaging Guided Transurethral Ultrasound Thermal TherapyJournal of Urology, 2007
- Contemporary Trends in Low Risk Prostate Cancer: Risk Assessment and TreatmentJournal of Urology, 2007
- Design and characterization of a digital image acquisition system for whole-specimen breast histopathologyPhysics in Medicine & Biology, 2006
- Control of thermal therapies with moving power deposition fieldPhysics in Medicine & Biology, 2006
- Analysis of factors important for transurethral ultrasound prostate heating using MR temperature feedbackPhysics in Medicine & Biology, 2006
- Method for MRI-guided conformal thermal therapy of prostate with planar transurethral ultrasound heating applicatorsPhysics in Medicine & Biology, 2005
- Closed-loop feedback control of phased-array microwave heating using thermal measurements from magnetic resonance imagingConcepts in Magnetic Resonance, 2002
- Prostate Cancer: MR Imaging and Thermometry during Microwave Thermal Ablation-Initial ExperienceRadiology, 2000
- Principal warps: thin-plate splines and the decomposition of deformationsPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1989