Effect of patient support pads on image quality and dose in fluoroscopy
- 1 March 1997
- journal article
- Published by Wiley in Medical Physics
- Vol. 24 (3) , 377-382
- https://doi.org/10.1118/1.597906
Abstract
An investigation was performed of the changes in image quality and patient dose as a result of increasing filtration for fluoroscopy performed under automatic brightness control. Filtration was added either at the x-ray tube housing (i.e., scatter-free geometry) or adjacent to a tissue equivalent phantom simulating the patient (i.e., with-scatter geometry). Patient doses were expressed in terms of the total energy imparted to patients simulated by either a 10 cm (i.e., pediatric) or 20 cm (i.e., adult) acrylic phantoms. Changes in image quality were determined by measuring the relative visibility of circular disks in a Leeds Test Object 10 contrast-detail phantom. In the scatter-free geometry, the addition of 4 mm Al filtration reduced the energy imparted by 27% (10 cm phantom) and 20% (20 cm phantom). In the with-scatter geometry, the corresponding reductions in energy imparted were 17% and 9% for the 10 and 20 cm phantoms, respectively. The visibility of low contrast disks generally decreased as the thickness of the added aluminum increased but the location of the added Al (i.e., with-scatter or scatter-free geometry) had no significant effect on the resultant image quality. These results demonstrate that the use of patient support pads with a thickness of approximately 4 mm Al will generally have an adverse impact on fluoroscopic image quality and result in modest reductions (approximately 10%) of adult patient doses.Keywords
This publication has 11 references indexed in Scilit:
- Evaluation of image quality in fluoroscopy by measurements and Monte Carlo calculationsPhysics in Medicine & Biology, 1995
- Semiempirical model for generating tungsten target x‐ray spectraMedical Physics, 1991
- Comparison of performance characteristics of conventional and K-edge filters in general diagnostic radiologyPhysics in Medicine & Biology, 1989
- The influence of tube filtration and potential on patient dose during X-ray examinationsPhysics in Medicine & Biology, 1988
- Filter materials for dose reduction in screen-film radiographyPhysics in Medicine & Biology, 1986
- A Set of X-Ray Test Objects for Quality Control in Television FluoroscopyThe British Journal of Radiology, 1985
- The measurement of energy imparted to patients during diagnostic x-ray examinations using the Diamentor exposure-area product meterPhysics in Medicine & Biology, 1984
- Energy imparted to water slabs by photons in the energy range 5-300 keV. Calculations using a Monte Carlo photon transport modelPhysics in Medicine & Biology, 1984
- “Burns” Occurring During Lengthy Surgical ProceduresJournal of Clinical Engineering, 1980