Analysis and compensation of eddy currents in balanced SSFP
Open Access
- 20 June 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 54 (1) , 129-137
- https://doi.org/10.1002/mrm.20527
Abstract
Balanced steady‐state free precession (SSFP) completely compensates for all gradients within each repetition time (TR), and is thus very sensitive to any magnetic field imperfection that disturbs the perfectly balanced acquisition scheme. It is demonstrated that balanced SSFP is especially sensitive to changing eddy currents that are induced by stepwise changing phase‐encoding (PE) gradients. In contrast to the linear k‐space trajectory, which has small variations between consecutive encoding steps, other encoding schemes (e.g., centric, random, or segmented orderings) exhibit significant jumps in k‐space between adjacent PE steps, and consequently induce rapidly changing eddy currents. The resulting disturbances induce significant image artifacts, such that compensation strategies are essential when nonlinear PE schemes are applied. Although direct annihilation of the induced eddy currents by additional, opposing magnetic fields has been investigated, it is limited by uncertainty regarding the time evolution of induced eddy currents. A generic (and thus system‐unrelated) compensation strategy is proposed that consists of “pairing” of consecutive PE steps. Another approach is based on partial dephasing along the slice direction that annihilates eddy‐current‐induced signal oscillations. Both pairing of the PE steps and “through‐slice equilibration” are easy to implement and allow the use of arbitrary k‐space trajectories for balanced SSFP. Magn Reson Med 54:129–137, 2005.Keywords
This publication has 9 references indexed in Scilit:
- View ordering for magnetization prepared steady state free precession acquisition: Application in contrast‐enhanced MR angiographyMagnetic Resonance in Medicine, 2004
- Accelerating cardiac cine 3D imaging using k‐t BLASTMagnetic Resonance in Medicine, 2004
- Is TrueFISP a gradient‐echo or a spin‐echo sequence?Magnetic Resonance in Medicine, 2003
- Single‐breathhold 3D‐trueFISP cine cardiac imagingMagnetic Resonance in Medicine, 2002
- A programmable pre‐emphasis systemMagnetic Resonance in Medicine, 1998
- On the stationary states in gradient echo imagingJournal of Magnetic Resonance Imaging, 1996
- Spoiling of transverse magnetization in steady‐state sequencesMagnetic Resonance in Medicine, 1991
- Motion‐insensitive, steady‐state free precession imagingMagnetic Resonance in Medicine, 1990
- Rapid Fourier imaging using steady‐state free precessionMagnetic Resonance in Medicine, 1987