Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Open Access
- 1 August 2004
- journal article
- Published by American Diabetes Association in Diabetes Care
- Vol. 27 (8) , 1922-1928
- https://doi.org/10.2337/diacare.27.8.1922
Abstract
OBJECTIVE—The objective of this study was to introduce continuous glucose–error grid analysis (CG-EGA) as a method of evaluating the accuracy of continuous glucose-monitoring sensors in terms of both accurate blood glucose (BG) values and accurate direction and rate of BG fluctuations and to illustrate the application of CG-EGA with data from the TheraSense Freestyle Navigator. RESEARCH DESIGN AND METHODS—We approach the design of CG-EGA from the understanding that continuous glucose sensors (CGSs) allow the observation of BG fluctuations as a process in time. We account for specifics of process characterization (location, speed, and direction) and for biological limitations of the observed processes (time lags associated with interstitial sensors). CG-EGA includes two interacting components: 1) point–error grid analysis (P-EGA) evaluates the sensor’s accuracy in terms of correct presentation of BG values and 2) rate–error grid analysis (R-EGA) assesses the sensor’s ability to capture the direction and rate of BG fluctuations. RESULTS—CG-EGA revealed that the accuracy of the Navigator, measured as a percentage of accurate readings plus benign errors, was significantly different at hypoglycemia (73.5%), euglycemia (99%), and hyperglycemia (95.4%). Failure to detect hypoglycemia was the most common error. The point accuracy of the Navigator was relatively stable over a wide range of BG rates of change, and its rate accuracy decreased significantly at high BG levels. CONCLUSIONS—Traditional self-monitoring of BG device evaluation methods fail to capture the important temporal characteristics of the continuous glucose-monitoring process. CG-EGA addresses this problem, thus providing a comprehensive assessment of sensor accuracy that appears to be a useful adjunct to other CGS performance measures.Keywords
This publication has 13 references indexed in Scilit:
- The Need for Separate Performance Goals for Glucose Sensors in the Hypoglycemic, Normoglycemic, and Hyperglycemic RangesDiabetes Care, 2004
- Timing of Changes in Interstitial and Venous Blood Glucose Measured With a Continuous Subcutaneous Glucose SensorDiabetes, 2003
- A Continuous Glucose Sensor Based on Wired Enzyme™ Technology - Results from a 3-Day Trial in Patients with Type 1 DiabetesDiabetes Technology & Therapeutics, 2003
- A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.Diabetes Care, 2000
- Performance Evaluation of the MiniMed® Continuous Glucose Monitoring System During Patient Home UseDiabetes Technology & Therapeutics, 2000
- Reservations on the Use of Error Grid Analysis for the Validation of Blood Glucose AssaysDiabetes Care, 1997
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- Accuracy of Perceiving Blood Glucose in IDDMDiabetes Care, 1985
- Amitriptyline Treatment of Painful Diabetic Neuropathy: An Inadvertent Single-Patient Clinical TrialDiabetes Care, 1985
- On Information and SufficiencyThe Annals of Mathematical Statistics, 1951