Application of heavy-ion CT
- 2 August 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (16) , 4073-4081
- https://doi.org/10.1088/0031-9155/51/16/013
Abstract
In treatment planning for hadron therapy, information about the relative stopping power in a patient's body is used to calculate the range of incident ions. This information is obtained from computed tomography (CT) images using a conversion table from x-ray CT numbers into stopping powers relative to the stopping power of water. In treatment planning at the National Institute of Radiological Sciences (NIRS), the conversion table has been created based on the polybinary tissue model. However, it has not been fully verified that the model is accurate enough for use in real animal tissues. In order to irradiate heavy ions more precisely in radiotherapy, we have to evaluate the accuracy of the polybinary tissue calibration in animal tissues. We have measured animal tissue samples with a heavy-ion CT (HICT) and an x-ray CT. The x-ray CT image was converted to an image of relative stopping power by using the table derived from the polybinary tissue calibration (polybinary-tissue-model CT (PTCT) image). On the other hand, with HICT, the two-dimensional distribution of relative stopping power can be obtained directly. A comparison between PTCT and HICT images enabled us to verify the accuracy of the conversion table derived from the polybinary tissue calibration. Consequently, it was found that the agreement between the relative stopping powers of PTCT and HICT is 1.6% for fat, muscle and bone.Keywords
This publication has 14 references indexed in Scilit:
- Measurement of electron density distribution using heavy ion CTNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- A CT calibration method based on the polybinary tissue model for radiotherapy treatment planningPhysics in Medicine & Biology, 2003
- Relation between carbon ion ranges and x‐ray CT numbersMedical Physics, 2001
- Relationship between CT number and electron density, scatter angle and nuclear reaction for hadron-therapy treatment planningPhysics in Medicine & Biology, 1998
- The precision of proton range calculations in proton radiotherapy treatment planning: experimental verification of the relation between CT-HU and proton stopping powerPhysics in Medicine & Biology, 1998
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996
- An electron density calibration phantom for CT-based treatment planning computersMedical Physics, 1992
- Possible use of proton CT as a means of density normalization in the PIXE semi-microprobe analysisNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1984
- Computed tomography using proton energy lossPhysics in Medicine & Biology, 1981
- Axial Scanning with 900 MeV Alpha ParticlesIEEE Transactions on Nuclear Science, 1975