On the applicability of the AAPM TG‐60/TG‐43 dose calculation formalism to intravascular line sources: Proposal for an adapted formalism

Abstract
Despite the widely recognized usefulness of the AAPM TG‐43 brachytherapy dose calculation formalism, a straightforward application of this approach to describe the dose distribution about intravascular line sources as proposed by TG‐60 may be difficult or even impossible, especially when these line sources emit low‐energy photons or beta particles. The causes of these limitations are investigated and illustrated by means of some numerical examples. In order to solve the observed limitations an adapted formalism is proposed, intended specifically for the description of the dose rate distribution about line sources but conceptually similar to the TG‐43/TG‐60 formalism. Several examples are presented to illustrate the usefulness of the proposed line source dose calculation formalism.