Commutability Assessment of Potential Reference Materials Using a Multicenter Split-Patient-Sample Between-Field-Methods (Twin-Study) Design: Study within the Framework of the Dutch Project “Calibration 2000”
Open Access
- 1 September 2002
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 48 (9) , 1520-1525
- https://doi.org/10.1093/clinchem/48.9.1520
Abstract
Background: The Dutch project “Calibration 2000” aims at harmonization of laboratory results via calibration by development of commutable, matrix-based, secondary reference materials. An alternative approach to the NCCLS EP14 protocol for studying commutability of reference materials is presented, the “twin-study design”, which in essence is a multicenter, split-patient-sample, between-field-methods protocol. Methods: The study consisted of the simultaneous analysis of fresh patient sera and potential reference materials (PRMs) for HDL-cholesterol (HDL-C) by 86 laboratories forming 43 laboratory couples. Six subgroups of method combinations were formed. The patient sera were selected and interchanged by each laboratory couple. The PRMs consisted of three types: C37, prepared according to the NCCLS C37 protocol; Fro, frozen selectively pooled human serum; and Lyo, which was the same serum pool as Fro but lyophilized in the presence of sucrose. All PRMs were provided in three HDL-C concentrations. The regression line residuals for the PRMs were normalized by expressing them as multiples of the state-of-the-art within laboratory SD (SDSA). In addition, the extra contribution of each PRM to the total measurement uncertainty, CVNetto, was calculated. Results: Averaged over the three PRM concentrations, 1.6% of the C37 residuals were outside the 3 SDSA limit. For the Fro and Lyo PRMs, these values were 2.4% and 11.1%. CVNetto values for C37, Fro, and Lyo were 2.9%, 4.3%, and 5.3%, respectively. Conclusions: The present twin-study design, as a practical alternative to the NCCLS EP14 protocol, is a viable way of studying commutability characteristics of PRMs. The study suggests that the C37 PRMs are the best candidates for a future reference material.Keywords
This publication has 7 references indexed in Scilit:
- Selection, Preparation, and Characterization of Commutable Frozen Human Serum Pools as Potential Secondary Reference Materials for Lipid and Apolipoprotein Measurements: Study within the Framework of the Dutch Project “Calibration 2000”Clinical Chemistry, 2002
- Implementation of Reference Systems in Laboratory MedicineClinical Chemistry, 2000
- A Reference Method Laboratory Network for Cholesterol: A Model for Standardization and Improvement of Clinical Laboratory MeasurementsClinical Chemistry, 2000
- The quest for comparability: Calibration 2000Accreditation and Quality Assurance, 2000
- Testing the accuracy of total cholesterol assays in an external quality-control program. Effect of adding sucrose to lyophilized control sera compared with use of fresh or frozen seraClinical Chemistry, 1995
- Accuracy in clinical chemistry--does anybody care?Clinical Chemistry, 1994
- Combi scheme: new combined internal/external quality-assessment scheme in The NetherlandsClinical Chemistry, 1991