Performance evaluation of a “dual‐side read” dedicated mammography computed radiography system
- 25 June 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (7) , 1843-1854
- https://doi.org/10.1118/1.1584045
Abstract
The image quality of a dedicated mammography computed radiography (CR) system was characterized. A unique feature of this system is that it collects image signals from both sides of the storage phosphor. Measurements of the modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were made. This work included improvements in our measurement methods to specifically account for the detrimental effects of system glare on the MTF and to accurately characterize the low-frequency NPS components. Image quality measurements were performed using a 25 kVp beam filtered with 2 mm Al and an exposure range of 1 to 100 mR (87 to 870 microGy). The DQE was found to decrease with increasing exposure due to an increased contribution of storage phosphor structure noise. The DQE of this system was compared to similar measurements made using a standard CR system. The dual-side read system demonstrated superior DQE compared to the standard system. The decrease in DQE with increasing exposure was more severe for the standard system than the dual-side read system. This finding suggests that the CR system noise was reduced for the dual-side read system compared to the standard system.Keywords
This publication has 32 references indexed in Scilit:
- A survey of clinical factors and patient dose in mammographyMedical Physics, 2001
- Regional improvement of signal-to-noise and contrast-to-noise ratios in dual-screen CR chest imaging-a phantom studyMedical Physics, 2001
- Comparison of Full-Field Digital Mammography with Screen-Film Mammography for Cancer Detection: Results of 4,945 Paired ExaminationsRadiology, 2001
- Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.Medical Physics, 2000
- A method for measuring the presampled MTF of digital radiographic systems using an edge test deviceMedical Physics, 1998
- Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammographyMedical Physics, 1997
- Physical evaluation of computed radiography as a mammographic X-ray imaging systemThe British Journal of Radiology, 1994
- Physical aspects of photostimulable phosphor computed radiographyThe British Journal of Radiology, 1993
- Epoxy resin based tissue substitutesThe British Journal of Radiology, 1977