Evaluation of the 4-hour RapID NF Plus method for identification of 345 gram-negative nonfermentative rods
- 1 May 1992
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 30 (5) , 1267-1270
- https://doi.org/10.1128/jcm.30.5.1267-1270.1992
Abstract
The ability of the RapID NF Plus system (Innovative Diagnostic Systems, Inc., Atlanta, Ga.) to identify 345 nonfermentative gram-negative rods was evaluated. Kits were inoculated with no. 1 McFarland suspensions, and reactions were interpreted after a 4-h incubation at 35 degrees C. Overall, the method correctly identified 311 strains (90.1%) without additional tests and 21 strains (6.1%) with additional tests, and 13 strains (3.8%) were misidentified. Five of 13 misidentified strains were Alcaligenes faecalis-Alcaligenes odorans misidentified as Alcaligenes xylosoxidans; however, all strains were xylose negative but nitrate positive and could have been A. faecalis group I-Alcaligenes piechaudii. The system does not differentiate between Pseudomonas fluorescens and Pseudomonas putida, and all Acinetobacter species are identified as Acetinobacter calcoaceticus. Additionally, no subspecies differentiation is made between A. xylosoxidans subsp. xylosoxidans and A. xylosoxidans subsp. denitrificans. All strains of the former Flavobacterium group IIb are identified as Flavobacterium indologenes-Flavobacterium gleum, and no species identification of the genus Methylobacterium is attempted. The system is easy to set up and interpret and provides an accurate commercial nonautomated method for same-day identification of gram-negative nonfermenters.Keywords
This publication has 10 references indexed in Scilit:
- Identification of clinical isolates of gram-negative nonfermentative bacteria by an automated cellular fatty acid identification systemJournal of Clinical Microbiology, 1991
- Preliminary evaluation of Biolog, a carbon source utilization method for bacterial identificationJournal of Clinical Microbiology, 1991
- Evaluation of Cathra system for identifying gram negative aerobic bacteria.Journal of Clinical Pathology, 1990
- Evaluation of the Titertek-NF system for identification of gram-negative nonfermentative and oxidase-positive fermentative bacteriaJournal of Clinical Microbiology, 1989
- Evaluation of PASCO MIC-ID system for identifying gram-negative bacilliJournal of Clinical Microbiology, 1987
- Critical evaluation of the AutoMicrobic system gram-negative identification card for identification of glucose-nonfermenting gram-negative rodsJournal of Clinical Microbiology, 1986
- autoSCAN-4 system for identification of gram-negative bacilliJournal of Clinical Microbiology, 1985
- Evaluation of the rapid NFT system for identification of gram-negative, nonfermenting rodsJournal of Clinical Microbiology, 1984
- Collaborative clinical evaluation of the Autobac IDX system for identification of gram-negative bacilliJournal of Clinical Microbiology, 1984
- Four methods for identification of gram-negative nonfermenting rods: organisms more commonly encountered in clinical specimensJournal of Clinical Microbiology, 1980