Temperature dependence of commercially available diode detectors
- 25 March 2002
- journal article
- Published by Wiley in Medical Physics
- Vol. 29 (4) , 622-630
- https://doi.org/10.1118/1.1461842
Abstract
Temperature dependence of commercially available n- and p-type diodes were studied experimentally under both high instantaneous dose rate (pulsed) and low dose rate (continuous) radiation. The sensitivity versus temperature was measured at SSD = 80 or 100 cm, 10 x 10 cm2, and 5 cm depth in a 30 x 30 x 30 cm3 water phantom between 10 degrees C and 35 degrees C. The response was linear for all the diode detectors. The temperature coefficient (or sensitivity variation with temperature, svwt) was dose rate independent for preirradiated diodes. They were (0.30 +/- 0.01)%/degrees C, (0.36 +/- 0.03)%/degrees C, and (0.29 +/- 0.08)%/degrees C for QED p-type, EDP p-type, and Isorad n-type diodes, respectively. The temperature coefficient for unirradiated n-type diodes was different under low dose rate [(0.16 to 0.45)%/degrees C, continuous, cobalt] and high instantaneous dose rate [(0.07 +/- 0.02)%/degrees C, pulsed radiation]. Moreover, the temperature coefficient varies among individual diodes. Similarly, the temperature coefficient for a special unirradiated QED p-type diode was different under low dose rate (0.34%/degrees C, cobalt) and high instantaneous dose rate [(0.26 +/- 0.01)%/degrees C, pulsed radiation]. Sufficient preirradiation can eliminate dose rate dependence of the temperature coefficient. On the contrary, preirradiation cannot eliminate dose rate dependence of the diode sensitivity itself.Keywords
This publication has 7 references indexed in Scilit:
- The thermal characteristics of different diodes on in vivo patient dosimetryMedical Physics, 2001
- In vivo dosimetry: Intercomparison between p‐type based and n‐type based diodes for the 16–25 MV energy rangeMedical Physics, 2000
- Linearity with dose rate of low resistivity p-type silicon semiconductor detectorsPhysics in Medicine & Biology, 1993
- Correlation between temperature and dose rate dependence of semiconductor response; influence of accumulated doseRadiotherapy and Oncology, 1990
- General specifications for silicon semiconductors for use in radiation dosimetryPhysics in Medicine & Biology, 1987
- Evaluation of temperature effects in p-type silicon detectorsPhysics in Medicine & Biology, 1986
- Effects of radiation damage on p-type silicon detectorsPhysics in Medicine & Biology, 1983