Evaluation of the MicroScan antimicrobial susceptibility system with the autoSCAN-4 automated reader
- 1 January 1986
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 23 (1) , 143-148
- https://doi.org/10.1128/jcm.23.1.143-148.1986
Abstract
The American MicroScan (American MicroScan, Mahwah, New Jersey, USA) identification and antimicrobial suseptibility system consists in part of an automated reading system (autoSCAN-4) with data management capabilities. We evaluated the system with 404 gram-negative and 170 gram-positive facultative anaerobic and aerobic bacteria. We compared MicroScan results read automatically and visually with each other and with the results obtained by the reference method (read visually). The overall agreement within .+-. 1 log2 dilution was 94.3% when the MicroScan system (read automatically) was compared with the reference method (read visually), 96.4% when MicroScan panels (read visually) were compared with reference panels, and 97.4% when the autoSCAN-4 automated reading was compared with the visual reading of the MicroScan panels. Total discrepancies (susceptibility interpretation category changes) for the MicroScan system compared with the reference method were 7%, with 6.2% considered a minor discrepancy. The autoSCAN-4 and the complete MicroScan system yielded accurate results compared with the reference method.This publication has 15 references indexed in Scilit:
- Evaluation of the AutoSCAN-3, a device for reading microdilution traysJournal of Clinical Microbiology, 1983
- Automated methods in microbiology: II. Identification and susceptibility testing.1983
- Automated reading of a microtitre plate: preliminary evaluation in antimicrobial susceptibility tests and Enterobacteriaceae identification.Journal of Clinical Pathology, 1983
- Evaluation of the MICUR system for quantitative antimicrobial susceptibility testing: a multiphasic comparison with reference methodsJournal of Clinical Microbiology, 1982
- Evaluation of the Sceptor microdilution antibiotic susceptibility testing system: a collaborative investigationJournal of Clinical Microbiology, 1981
- Reader error in determining minimal inhibitory concentrations with microdilution susceptibility test panelsJournal of Clinical Microbiology, 1981
- Evaluation of the sensititre microdilution antibiotic susceptibility system against recent clinical isolates: three-laboratory collaborative studyJournal of Clinical Microbiology, 1980
- Determination of antibiotic sensitivities by the Sensititre system.Journal of Clinical Pathology, 1978
- Evaluation of the Micro-Media System for Quantitative Antimicrobial Drug Susceptibility Testing: a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1978
- Evaluation of the Performance Parameters of a Prediluted, Quantitative Antibiotic Susceptibility Test DeviceAntimicrobial Agents and Chemotherapy, 1977