Eccentric and concentric isokinetic knee flexion and extension: a reliability study using the Cybex 6000 dynamometer.
Open Access
- 1 June 1996
- journal article
- research article
- Published by BMJ in British Journal of Sports Medicine
- Vol. 30 (2) , 156-160
- https://doi.org/10.1136/bjsm.30.2.156
Abstract
OBJECTIVE: To determine the reliability of the Cybex 6000 isokinetic dynamometer in measuring the knee muscle performance concentrically and eccentrically. METHODS: 18 male and 12 female subjects with no previous knee injuries, who had not previously undergone any isokinetic testing, were studied. The flexor and extensor muscles groups of both knees were tested at 60 degrees s-1 and 120 degrees s-1 with the continuous concentric-eccentric cycle testing protocol. Variables studied included peak torque, total work, and average power. The interclass correlation coefficient (ICC 2,1) was used to determine the reliability with P < 0.05. RESULTS: Peak torque showed significantly greater ICC than the total work and average power, with test-retest reliability ranging from 0.82 to 0.91 for peak torque, from 0.76 to 0.89 for total work, and 0.71 to 0.88 for average power. Average variability for the three variables studied ranged from 9% to 14%. The ICCs for the three variables studied were significantly greater at 120 degrees s-1. The knee extensor muscle group showed greater test-retest reliability, and the results of isokinetic testing in the concentric contraction mode were more reproducible. CONCLUSIONS: The Cybex 6000 isokinetic dynamometer shows high reliability in measuring isokinetic concentric and eccentric variables. Some fluctuation should be allowed when evaluating variations of muscle performance between tests.Keywords
This publication has 32 references indexed in Scilit:
- Reliability of eccentric isokinetic knee flexion and extension measurementsArchives of Physical Medicine and Rehabilitation, 1993
- The strength characteristics of internal and external rotator muscles in professional baseball pitchersThe American Journal of Sports Medicine, 1993
- Torque-velocity relationships for the knee extensors in women in their 3rd and 7th decadesEuropean Journal of Applied Physiology, 1990
- Comparison of Values Generated During Testing of the Knee Using the Cybex II Plus®and Biodex Model B-2000®Isokinetic DynamometersJournal of Orthopaedic & Sports Physical Therapy, 1989
- Reliability of an Isokinetic Test of Muscle Strength and EnduranceJournal of Orthopaedic & Sports Physical Therapy, 1989
- Reliability of a Reciprocal Test Protocol Performed on the Kinetic Communicator: An Isokinetic Test of Knee Extensor and Flexor StrengthJournal of Orthopaedic & Sports Physical Therapy, 1988
- Upper extremity range of motion and isokinetic strength of the internal and external shoulder rotators in major league baseball playersThe American Journal of Sports Medicine, 1988
- Comparison of Peak Torque Values of the Knee Flexor and Extensor Muscle Groups Using the Cybex II®and Lido 2.0®Isokinetic DynamometersJournal of Orthopaedic & Sports Physical Therapy, 1987
- The relationship of upper extremity strength to throwing speedThe American Journal of Sports Medicine, 1982
- VARIABILITY OF ISOKINETIC MEASURESMedicine & Science in Sports & Exercise, 1982