Detection of Insulation Failure by Gradual Reduction in Noninvasively Measured Electrogram Amplitudes
- 1 September 1986
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 9 (5) , 772-775
- https://doi.org/10.1111/j.1540-8159.1986.tb05427.x
Abstract
Insulation failure of pacemaker leads may give rise to functional defects such as under- or oversensing, loss of capture or muscle stimulation. An increased incidence of such failures has been reported for the bipolar polyurethane-insulated Medtronic model 6972 pacing lead. It is therefore important to identify techniques that will detect such insulation defects early. A case report is presented in which diminishing electrogram amplitude at follow-up predicted lead failure several months before becoming clinically manifest. This finding suggests that routine measurement of such an electrogram may be a useful tool in predicting impending lead malfunction.Keywords
This publication has 4 references indexed in Scilit:
- Sixteen Failures in a Single Model of Bipolar Polyurethane‐insulated Ventricular Pacing Lead: A 44‐Month ExperiencePacing and Clinical Electrophysiology, 1984
- Insulation Failure with Bipolar Polyurethane Pacing LeadsPacing and Clinical Electrophysiology, 1984
- False Inhibition of an Atrial Demand Pacemaker Caused by an Insulation Defect in a Polyurethane LeadPacing and Clinical Electrophysiology, 1983
- Polyurethane Leads: Facts and ControversyPacing and Clinical Electrophysiology, 1983