Sixteen Failures in a Single Model of Bipolar Polyurethane‐insulated Ventricular Pacing Lead: A 44‐Month Experience
- 1 May 1984
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 7 (3) , 389-394
- https://doi.org/10.1111/j.1540-8159.1984.tb04923.x
Abstract
During a 44-month period, 105 Medtronic model 6972-58 polyurethane-insulated, bipolar ventricular pacing leads were implanted, of which 16 failed clinically at a mean of 20.5 months post-implant. Pacing dysfunction presented as either intermittent or complete loss of sensing and/or capture and, rarely, oversensing . Thirteen of the 16 patients were asymptomatic, and problems were detected in 11 instances during routine telephonic surveillance. At the time of lead replacement, low or widely fluctuating lead resistance values and high current drain were uniformly observed. The single lead capable of total extraction revealed rupture of both internal and external insulation. A brief summary of the currently much-discussed "polyurethane controversy" is presented, and the issue of patient management is discussed.Keywords
This publication has 19 references indexed in Scilit:
- Self‐Inhibition of an AV Sequential Demand (DVI) Pulse Generator Due to Polyurethane Lead Insulation DisruptionPacing and Clinical Electrophysiology, 1983
- False Inhibition of an Atrial Demand Pacemaker Caused by an Insulation Defect in a Polyurethane LeadPacing and Clinical Electrophysiology, 1983
- Polyurethane LeadsPacing and Clinical Electrophysiology, 1983
- The Polyurethane ControversyPacing and Clinical Electrophysiology, 1983
- Polyurethane Leads: Facts and ControversyPacing and Clinical Electrophysiology, 1983
- Use of Synthetic Polymers for Biomedical ApplicationPacing and Clinical Electrophysiology, 1983
- STIMAREC Report of May‐June, 1982Pacing and Clinical Electrophysiology, 1982
- Autointerference of Demand Pulse GeneratorsPacing and Clinical Electrophysiology, 1981
- Comprehensive Cardiac Pacemaker Information System: Basis for a Regional Follow‐up NetworkPacing and Clinical Electrophysiology, 1978
- Scanning electron microscopy of surface irregularities and thrombogenesis of polyurethane and polyethylene coronary catheters.Circulation, 1976