Mass Output and Particle Size Distribution of Glucocorticosteroids Emitted from Different Inhalation Devices Depending on Various Inspiratory Parameters
- 1 March 2002
- journal article
- research article
- Published by Mary Ann Liebert Inc in Journal of Aerosol Medicine
- Vol. 15 (1) , 65-73
- https://doi.org/10.1089/08942680252908593
Abstract
Efficient inhalation therapy depends on successful delivery of the drug to the lung. The efficacy of drug delivery is not only influenced by the characteristics of the inhalation device, but also by the patient's handling of the device and by the inspiratory maneuver achieved through the device. We analyzed the output characteristics of three different chlorofluorocarbon (CFC)-free breath-actuated inhalers for inhaled glucocorticosteroids (BUD Turbohaler®, FP Diskus®/Accuhaler® and HFA-BDP Autohaler®, respectively). Mass output and particle size distribution of drug aerosol delivered by the inhalers were determined depending on different inhalation parameters in vitro using an Andersen cascade impactor. We found that, beside the peak inspiratory flow (PIF), other factors such as flow acceleration and inhalation volume also have significant effects on aerosol generation with respect to mass output and particle size distribution. Thus, these parameters should be taken into account when a suitable device for an individual patient is to be selected. The dependency on inspiratory parameters was most pronounced for the dry powder inhalers. The Turbohaler® showed by far the highest variance in particle output (fine particle fraction ranging from 3.4% to 22.1% of label claim), whereas the Diskus® was less dependent on variations in inhalation (10.6% to 18.5% of label claim). The most constant aerosol output was found for the Autohaler®, which also released the highest fine particle fraction (43.1% to 56.6% of label claim).Keywords
This publication has 17 references indexed in Scilit:
- In vitro Comparison of Three Salbutamol-Containing Multidose Dry Powder InhalersClinical Drug Investigation, 2000
- A Critical Comparison of the Dose Delivery Characteristics of Four Alternative Inhalation Devices Delivering Salbutamol: Pressurized Metered Dose Inhaler,DiskusInhaler,DiskhalerInhaler, andTurbuhalerInhalerJournal of Aerosol Medicine, 1999
- Design, validation and initial testing of the Electronic LungTM deviceJournal of Aerosol Science, 1998
- Influence of Flow Rate on Aerosol Particle Size Distributions from Pressurized and Breath-Actuated InhalersJournal of Aerosol Medicine, 1998
- Flow early in the inspiratory manoeuvre affects the aerosol particle size distribution from a TurbuhalerRespiratory Medicine, 1997
- Effect of Inhalation Flow Rate on the Dosing Characteristics of Dry Powder Inhaler (DPI) and Metered Dose Inhaler (MDI) ProductsJournal of Aerosol Medicine, 1996
- Dose emissions from marketed dry powder inhalersInternational Journal of Pharmaceutics, 1995
- A Novel Breath Actuated Device (Autohaler™) Consistently Actuates During the Early Phase of InspirationJournal of Aerosol Medicine, 1995
- A Model for the Effect of Inhalation Device Flow Resistance on the Peak Inspiratory Flow Rate and Its Application in Pharmaceutical TestingJournal of Aerosol Medicine, 1994
- Turbuhaler: A New Powder Inhaler for Administration of Drugs to the AirwaysPharmaceutical Research, 1988