The effects of intra‐fraction organ motion on the delivery of intensity‐modulated field with a multileaf collimator
- 25 June 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (7) , 1736-1746
- https://doi.org/10.1118/1.1578771
Abstract
Intensity‐modulated radiation therapy can be conveniently delivered with a multileaf collimator. With this method, the entire field is not delivered at once, but rather it is composed of many subfields defined by the leaf positions as a function of beam on time. At any given instant, only these subfields are delivered. During treatment, if the organ moves, part of the volume may move in or out of these subfields. Due to this interplay between organ motion and leaf motion the delivered dose may be different from what was planned. In this work, we present a method that calculates the effects of organ motion on delivered dose. The direction of organ motion may be parallel or perpendicular to the leaf motion, and the effect can be calculated for a single fraction or for multiple fractions. Three breast patients and four lung patients were included in this study, with the amplitude of the organ motion varying from to and the period varying from 4 to 8 seconds. Calculations were made for these patients with and without organ motion, and results were examined in terms of isodose distribution and dose volume histograms. Each calculation was repeated ten times in order to estimate the statistical uncertainties. For selected patients, calculations were also made with conventional treatment technique. The effects of organ motion on conventional techniques were compared relative to that on IMRT techniques. For breast treatment, the effect of organ motion primarily broadened the penumbra at the posterior field edge. The dose in the rest of the treatment volume was not significantly affected. For lung treatment, the effect also broadened the penumbra and degraded the coverage of the planning target volume (PTV). However, the coverage of the clinical target volume (CTV) was not much affected, provided the PTV margin was adequate. The same effects were observed for both IMRT and conventional treatment techniques. For the IMRT technique, the standard deviations of ten samples of a 30‐fraction calculation were very small for all patients, implying that over a typical treatment course of 30 fractions, the delivered dose was very close to the expected value. Hence, under typical clinical conditions, the effect of organ motion on delivered dose can be calculated without considering the interplay between the organ motion and the leaf motion. It can be calculated as the weighted average of the dose distribution without organ motion with the distribution of organ motion. Since the effects of organ motion on dose were comparable for both IMRT and conventional techniques, the PTV margin should remain the same for both techniques.Keywords
This publication has 20 references indexed in Scilit:
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002
- A simplified intensity modulated radiation therapy technique for the breastMedical Physics, 2002
- Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy systemJournal of Applied Clinical Medical Physics, 2001
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer?Radiotherapy and Oncology, 1998
- A gradient inverse planning algorithm with dose‐volume constraintsMedical Physics, 1998
- An investigation of tomotherapy beam deliveryMedical Physics, 1997
- Respiration gated radiotherapy treatment: a technical studyPhysics in Medicine & Biology, 1996
- Dose calculation for photon beams with intensity modulation generated by dynamic jaw or multileaf collimationsMedical Physics, 1994
- Analysis of movement of intrathoracic neoplasms using ultrafast computerized tomographyInternational Journal of Radiation Oncology*Biology*Physics, 1990