Work of Breathing During Constant- and Variable-Flow Nasal Continuous Positive Airway Pressure in Preterm Neonates
- 1 September 2001
- journal article
- clinical trial
- Published by American Academy of Pediatrics (AAP) in Pediatrics
- Vol. 108 (3) , 682-685
- https://doi.org/10.1542/peds.108.3.682
Abstract
Background. Constant-flow nasal continuous positive airway pressure (NCPAP) often is used in preterm neonates to recruit and maintain lung volume. Physical model studies indicate that a variable-flow NCPAP device provides more stable volume recruitment with less imposed work of breathing (WOB). Although superior lung recruitment with variable-flow NCPAP has been demonstrated in preterm neonates, corroborating WOB data are lacking. Objective. To measure and compare WOB associated with the use of variable-flow versus constant-flow NCPAP in preterm neonates. Methods. Twenty-four preterm infants who were receiving constant-flow NCPAP (means, SD) and had birth weight of 1024 ± 253 g, gestational age of 28 ± 1.7 weeks, age of 14 ± 13 days, and Fio 2 of 0.3 ± 0.1 were studied. Variable-flow and constant-flow NCPAP were applied in random order. We measured changes in lung volume and tidal ventilation (VT) by DC-coupled/calibrated respiratory inductance plethysmography as well as esophageal pressures at NCPAP of 8, 6, 4, and 0 cm H2O. Inspiratory WOB (WOBI) and lung compliance were calculated from the esophageal pressure and VT data using standard methods. WOB was divided by VT to standardize the results. Results. WOBI decreased at all CPAP levels with variable-flow NCPAP, with a maximal decrease at 4 cm H2O. WOBI increased at all CPAP levels with constant-flow CPAP. Lung compliance increased at all NCPAP levels with variable-flow, with a relative decrease at 8 cm H2O, whereas it increased only at 8 cm H2O with constant-flow NCPAP. Compared with constant-flow NCPAP, WOBI was 13% to 29% lower with variable-flow NCPAP. Conclusion. WOBI is decreased with variable-flow NCPAP compared with constant-flow NCPAP. The increase in WOBI with constant-flow NCPAP indicates the presence of appreciable imposed WOB with this device. Our study, performed in neonates with little lung disease, indicates the possibility of lung overdistention at CPAP of 6 to 8 cm H2O with the variable-flow device. Further study is necessary to determine the efficacy of variable-flow NCPAP in neonates with significant lung disease and its use over extended periods of time.continuous-flow and variable-flow NCPAP, work of breathing, premature neonates, lung compliance.Keywords
This publication has 11 references indexed in Scilit:
- Lung Recruitment and Breathing Pattern During Variable Versus Continuous Flow Nasal Continuous Positive Airway Pressure in Premature Infants: An Evaluation of Three DevicesPediatrics, 2001
- Do Clinical Markers of Barotrauma and Oxygen Toxicity Explain Interhospital Variation in Rates of Chronic Lung Disease?Pediatrics, 2000
- Lung resistance and elastance in spontaneously breathing preterm infants: effects of breathing pattern and demographicsJournal of Applied Physiology, 2000
- Randomized Controlled Trial of Two Methods of Nasal CPAP(NCPAP): Flow Driver Vs Conventional NCPAPPediatric Research, 1999
- A Crossover Trial of Infant Flow (IF) Continuous Positive Airway Pressure (CPAP) Versus Nasopharyngeal (NP) CPAP in the Extubation of Babies ≤ 1250 Grams BirthweightPediatric Research, 1999
- Decreased imposed work with a new nasal continuous positive airway pressure devicePediatric Pulmonology, 1996
- Nasal continuous positive airway pressure: experiences with a new technical approachActa Paediatrica, 1993
- Effect of nasal CPAP on thoracoabdominal motion in neonates with respiratory insufficiencyPediatric Pulmonology, 1991
- A new device for administration of nasal continuous positive airway pressure in the newbornCritical Care Medicine, 1988
- Measurement of Tidal Volume during High Frequency Ventilation by Impedance PlethysmographyPediatric Research, 1988