Technical considerations of dual-energy X-ray absorptiometry-based bone mineral measurements for pediatric studies
Open Access
- 1 December 1995
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 10 (12) , 1998-2004
- https://doi.org/10.1002/jbmr.5650101220
Abstract
Dual-energy X-ray absorptiometry (DXA) measurements have been shown to provide useful information on bone mineral status in young pediatric subjects. The purpose of this study was to challenge this system under various conditions to determine the clinical and experimental parameters that may be encountered which could interfere with DXA-based bone mineral content (BMC) and bone mineral density (BMD) measurements. Variations in data acquisition, including the covering of step phantom (external calibration standard) with a cotton blanket or partial exclusion of step phantom in the scan field, tissue freezing, or the presence of small nonmetallic objects, did not significantly alter DXA BMC or BMD measurements. By contrast, the presence of movement artifact, radiographic contrast media, and nonmetallic orthopedic casts significantly interfered with DXA BMC and BMD measurements. Variability in operator-dependent analysis of DXA scans occurred with regional analysis of whole body scans for DXA BMC and BMD measurements (average coefficient of variation was 2.9% and 1%, respectively, depending on the region analyzed) but did not affect the total (whole body) result. A minor adjustment in the manual delineation of the step phantom during data analysis may result in almost a 30% difference in DXA BMC and BMD. We conclude that movement artifact, radiographic contrast media, nonmetallic orthopedic cast, and variations in operator-dependent data analysis may interfere with DXA BMC and BMD measurement in young pediatric subjects. Therefore, appropriate care should be taken to reduce or eliminate such interference.Keywords
Funding Information
- NICHD (N01-HD-1-3126)
- University of Tennessee Medical Group Research Grant, The University of Tennessee—Memphis, Clinical Research Center (USPHS RR00211–29)
- Hologic, Inc.
This publication has 9 references indexed in Scilit:
- Validation of accuracy and precision of dual energy X-ray absorptiometry for infantsJournal of Bone and Mineral Research, 1995
- Accuracy of dual-energy x-ray absorptiometry for body-composition measurements in childrenThe American Journal of Clinical Nutrition, 1994
- Validation and application of dual-energy x-ray absorptiometry to measure bone mass and body composition in small infantsThe American Journal of Clinical Nutrition, 1993
- RADIOLOGIC DIAGNOSIS OF OSTEOPOROSISRadiologic Clinics of North America, 1993
- Bone mass measurement by DXA: Influence of analysis procedures and interunit variationJournal of Bone and Mineral Research, 1992
- Diagnostic and therapeutic technology assessment. Measurement of bone density with dual-energy X-ray absorptiometry (DEXA)Published by American Medical Association (AMA) ,1992
- Clinical Use of Bone DensitometryNew England Journal of Medicine, 1991
- Dual-energy x-ray absorptiometry for total-body and regional bone-mineral and soft-tissue compositionThe American Journal of Clinical Nutrition, 1990
- Immobilization and boneCalcified Tissue International, 1983