Ambient Mass Spectrometry with a Handheld Mass Spectrometer at High Pressure
- 15 September 2007
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 79 (20) , 7734-7739
- https://doi.org/10.1021/ac071114x
Abstract
The first coupling of atmospheric pressure ionization methods, electrospray ionization (ESI) and desorption electrospray ionization (DESI), to a miniature hand-held mass spectrometer is reported. The instrument employs a rectilinear ion trap (RIT) mass analyzer and is battery-operated, hand-portable, and rugged (total system: 10 kg, 0.014 m3, 75 W power consumption). The mass spectrometer was fitted with an atmospheric inlet, consisting of a 10 cm × 127 μm inner diameter stainless steel capillary tube which was used to introduce gas into the vacuum chamber at 13 mL/min. The operating pressure was 15 mTorr. Ions, generated by the atmospheric pressure ion source, were directed by the inlet along the axis of the ion trap, entering through an aperture in the dc-biased end plate, which was also operated as an ion gate. ESI and DESI sources were used to generate ions; ESI-MS analysis of an aqueous mixture of drugs yielded detection limits in the low parts-per-billion range. Signal response was linear over more than 3 orders of magnitude. Tandem mass spectrometry experiments were used to identify components of this mixture. ESI was also applied to the analysis of peptides and in this case multiply charged species were observed for compounds of molecular weight up to 1200 Da. Cocaine samples deposited or already present on different surfaces, including currency, were rapidly analyzed in situ by DESI. A geometry-independent version of the DESI ion source was also coupled to the miniature mass spectrometer. These results demonstrate that atmospheric pressure ionization can be implemented on simple portable mass spectrometry systems.Keywords
This publication has 20 references indexed in Scilit:
- Tissue Imaging at Atmospheric Pressure Using Desorption Electrospray Ionization (DESI) Mass SpectrometryAngewandte Chemie International Edition in English, 2006
- Generation and detection of multiply-charged peptides and proteins by matrix-assisted laser desorption electrospray ionization (MALDESI) fourier transform ion cyclotron resonance mass spectrometryPublished by American Chemical Society (ACS) ,2006
- Detection and quantification of chemical plumes using a portable underwater membrane introduction mass spectrometerTrAC Trends in Analytical Chemistry, 2006
- Microfabricated Quadrupole Ion Trap for Mass Spectrometer ApplicationsPhysical Review Letters, 2006
- Electrospray‐assisted laser desorption/ionization mass spectrometry for direct ambient analysis of solidsRapid Communications in Mass Spectrometry, 2005
- Ambient mass spectrometry using desorption electrospray ionization (DESI): instrumentation, mechanisms and applications in forensics, chemistry, and biologyJournal of Mass Spectrometry, 2005
- Rectilinear Ion Trap: Concepts, Calculations, and Analytical Performance of a New Mass AnalyzerAnalytical Chemistry, 2004
- Miniature Cylindrical Ion Trap Mass SpectrometerAnalytical Chemistry, 2002
- Imaging mass spectrometry: A new technology for the analysis of protein expression in mammalian tissuesNature Medicine, 2001
- Miniature mass analyzersJournal of Mass Spectrometry, 2000