Optimized T1‐weighted contrast for single‐slab 3D turbo spin‐echo imaging with long echo trains: Application to whole‐brain imaging
Open Access
- 29 October 2007
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 58 (5) , 982-992
- https://doi.org/10.1002/mrm.21386
Abstract
T1-weighted contrast is conventionally obtained using multislice two-dimensional (2D) spin-echo (SE) imaging. Achieving isotropic, high spatial resolution is problematic with conventional methods due to a long acquisition time, imperfect slice profiles, or high-energy deposition. Single-slab 3D SE imaging was recently developed employing long echo trains with variable low flip angles to address these problems. However, long echo trains may yield suboptimal T1-weighted contrast, since T2 weighting of the signals tends to develop along the echo train. Image blurring may also occur if high spatial frequency signals are acquired with low signal intensity. The purpose of this work was to develop an optimized T1-weighted version of single-slab 3D SE imaging with long echo trains. Refocusing flip angles were calculated based on a tissue-specific prescribed signal evolution. Spatially nonselective excitation was used, followed by half-Fourier acquisition in the in-plane phase encoding (PE) direction. Restore radio frequency (RF) pulses were applied at the end of the echo train to optimize T1-weighted contrast. Imaging parameters were optimized by using Bloch equation simulation, and imaging studies of healthy subjects were performed to investigate the feasibility of whole-brain imaging with isotropic, high spatial resolution. The proposed technique permitted highly-efficient T1-weighted 3D SE imaging of the brain. Magn Reson Med 58:982–992, 2007.Keywords
This publication has 19 references indexed in Scilit:
- Actual flip‐angle imaging in the pulsed steady state: A method for rapid three‐dimensional mapping of the transmitted radiofrequency fieldMagnetic Resonance in Medicine, 2006
- Factors influencing flip angle mapping in MRI: RF pulse shape, slice‐select gradients, off‐resonance excitation, and B0 inhomogeneitiesMagnetic Resonance in Medicine, 2006
- Fast spin echo sequences with very long echo trains: Design of variable refocusing flip angle schedules and generation of clinical T2 contrastMagnetic Resonance in Medicine, 2006
- Experimental analysis of parallel excitation using dedicated coil setups and simultaneous RF transmission on multiple channelsMagnetic Resonance in Medicine, 2005
- Multiecho sequences with variable refocusing flip angles: Optimization of signal behavior using smooth transitions between pseudo steady states (TRAPS)Magnetic Resonance in Medicine, 2003
- HyperechoesMagnetic Resonance in Medicine, 2001
- Three‐dimensional fast spin‐echo imaging: Pulse sequence and in vivo image evaluationJournal of Magnetic Resonance Imaging, 1993
- T2‐weighted thin‐section imaging with the multislab three‐dimensional RARE techniqueJournal of Magnetic Resonance Imaging, 1991
- Contrast manipulation and artifact assessment of 2D and 3D RARE sequencesMagnetic Resonance Imaging, 1990
- RARE imaging: A fast imaging method for clinical MRMagnetic Resonance in Medicine, 1986