Evaluation of the Abbott m 2000 RealTi m e Human Immunodeficiency Virus Type 1 (HIV-1) Assay for HIV Load Monitoring in South Africa Compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 Assays
- 1 July 2009
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 47 (7) , 2209-2217
- https://doi.org/10.1128/jcm.01761-08
Abstract
The implementation of antiretroviral therapy demands the need for increased access to viral load (VL) monitoring. Newer real-time VL testing technologies are faster and have larger dynamic ranges and fully automated extraction to benefit higher throughputs in resource-poor environments. The Abbott RealTime human immunodeficiency virus type 1 (HIV-1) assay was evaluated as a new option for testing for HIV-1 subtype C in South Africa, and its performance was compared to the performance of existing assays (the Cobas AmpliPrep-Cobas TaqMan HIV-1, version 1, assay; the AmpliPrep-Cobas Monitor standard HIV-1 assay; and the NucliSENS EasyQ-EasyMag HIV-1 assay) in a high-throughput laboratory. The total precision of the RealTime HIV-1 assay was acceptable over all viral load ranges. This assay compared most favorably with the Cobas AmpliPrep-Cobas TaqMan HIV-1 assay (R2 = 0.904), with a low standard deviation of difference being detected (0.323 copies/ml). The bias against comparator assays ranged from −0.001 copies/ml to −0.228 copies/ml. Variability in the reporting of VLs for a 20-member subtype panel compared to the variability of other assays was noted with subtypes G and CRF02-AG. The RealTime HIV-1 assay can test 93 samples per day with minimal manual preparation, less staff, and the minimization of contamination through automation. This assay is suitable for HIV-1 subtype C VL quantification in South Africa.Keywords
This publication has 44 references indexed in Scilit:
- Comparative Evaluation of the Automated Roche TaqMan Real-Time Quantitative Human Immunodeficiency Virus Type 1 RNA PCR Assay and the Roche AMPLICOR Version 1.5 Conventional PCR AssayJournal of Clinical Microbiology, 2007
- Human Immunodeficiency Virus Type 1 (HIV-1) Plasma Load Discrepancies between the Roche COBAS AMPLICOR HIV-1 MONITOR Version 1.5 and the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 AssaysJournal of Clinical Microbiology, 2007
- Thermodynamically modulated partially double-stranded linear DNA probe design for homogeneous real-time PCRNucleic Acids Research, 2007
- Evaluation of the Performance of the Automated NucliSENS easyMAG and EasyQ Systems versus the Roche AmpliPrep-AMPLICOR Combination for High-Throughput Monitoring of Human Immunodeficiency Virus LoadJournal of Clinical Microbiology, 2007
- Characteristics of the m 2000 Automated Sample Preparation and Multiplex Real-Time PCR System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeaeJournal of Clinical Microbiology, 2007
- Comparison of the Roche COBAS Amplicor™ Monitor, Roche COBAS Ampliprep™/COBAS Taqman™ and Abbott RealTime™ Test assays for quantification of hepatitis C virus and HIV RNAcclm, 2007
- Multicenter Performance Evaluation of a New TaqMan PCR Assay for Monitoring Human Immunodeficiency Virus RNA LoadJournal of Clinical Microbiology, 2006
- HIV-1 Viral Load Assays for Resource-Limited SettingsPLoS Medicine, 2006
- Comparative Evaluation of the Performance of the Abbott Real-Time Human Immunodeficiency Virus Type 1 (HIV-1) Assay for Measurement of HIV-1 Plasma Viral Load following Automated Specimen PreparationJournal of Clinical Microbiology, 2006
- Molecular Beacons: Probes that Fluoresce upon HybridizationNature Biotechnology, 1996