Selection of anti-scatter grids for different imaging tasks: the advantage of low atomic number cover and interspace materials
- 1 December 1993
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 66 (792) , 1151-1163
- https://doi.org/10.1259/0007-1285-66-792-1151
Abstract
A Monte Carlo computer program has been developed for the study of anti-scatter grids used in diagnostic radiology. The program estimates the scatter from soft tissue phantoms representative of either adult or paediatric examinations and uses dose increase, signal-to-noise ratio improvement and contrast improvement factors to study grid performance. It has been used to quantify the advantage of replacing grids with aluminium covers and interspaces by grids using materials of low atomic number for these components. Two approaches are used. First, the aluminium and low atomic number alternatives are compared for five grid ratios at fixed strip density and width and for tube potentials of 50, 70, 100 and 150 kV. Second, 44 commercially available grids are compared for three different imaging situations (lumbar spine, chest and paediatric). The results demonstrate that grids made with carbon fibre cover and cotton fibre interspace result in greater improvements in contrast and signal-to-noise ratio, and lower dose increase factors, than do grids made with aluminium. The dose reduction varies with irradiation conditions and is generally larger at lower tube potentials, higher grid ratios and lower strip densities. A typical reduction in mean absorbed dose in the patient is 30% in an adult lumbar spine (AP view) at 70 kV with a grid with 36 strips per centimetre and ratio 12.Keywords
This publication has 26 references indexed in Scilit:
- Monte Carlo study of grid performance in diagnostic radiology: task dependent optimization for screen–film imagingThe British Journal of Radiology, 1994
- Monte Carlo study of grid performance in diagnostic radiology: factors which affect the selection of tube potential and grid ratioThe British Journal of Radiology, 1993
- Calculation of dose and contrast for two mammographic gridsPhysics in Medicine & Biology, 1992
- The use of carbon fibre material in table tops, cassette fronts and grid covers: magnitude of possible dose reductionThe British Journal of Radiology, 1986
- Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinationsMedical Physics, 1984
- A Monte Carlo program for photon transport using analogue sampling of scattering angle in coherent and incoherent scattering processesComputer Programs in Biomedicine, 1983
- A method for testing scattered radiation gridsMedical Physics, 1980
- Quality and Choice of Potter Bucky Grids Parts IV and VActa Radiologica, 1961