Accuracy of the Ysi Stat Plus Analyzer for Glucose and Lactate
- 1 March 1995
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 28 (4) , 581-592
- https://doi.org/10.1080/00032719508001119
Abstract
The YSI model 2300 Stat Plus analyzer accurately measures glucose in aqueous solutions with an overall standard error of prediction (SEP) of 0.28 mM and a mean percent error (MPE) of 1.79 %. The absolute and relative errors for glucose measurements are concentration dependent. Initial measurements demonstrate the need to periodically monitor the accuarcy of this analyzer to avoid systematic errors. Lactate is measured with an overall SEP of 0.18 mM and MPE of 5.10 %. Analysis of the residuals indicates a slight negative bias of -0.07 mM for lactate.Keywords
This publication has 2 references indexed in Scilit:
- Temperature-Insensitive Near-Infrared Spectroscopic Measurement of Glucose in Aqueous SolutionsApplied Spectroscopy, 1994
- Noninvasive Blood Glucose Assay by Near-Infrared Diffuse Reflectance Spectroscopy of the Human Inner LipApplied Spectroscopy, 1993