Respiratory gating of cardiac PET data in list-mode acquisition
- 17 January 2006
- journal article
- research article
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 33 (5) , 584-588
- https://doi.org/10.1007/s00259-005-0031-0
Abstract
Purpose Respiratory motion has been identified as a source of artefacts in most medical imaging modalities. This paper reports on respiratory gating as a means to eliminate motion-related inaccuracies in PET imaging. Methods Respiratory gating was implemented in list mode with physiological signal recorded every millisecond together with the PET data. Respiration was monitored with an inductive respiration monitor using an elasticised belt around the patient’s chest. Simultaneous ECG gating can be maintained independently by encoding ECG trigger signal into the list-mode data. Respiratory gating is performed in an off-line workstation with gating parameters defined retrospectively. The technique was applied on a preliminary set of patient data with C15O. Results Motion was visually observed in the cine displays of the sagittal and coronal views of the reconstructed respiratory gated images. Significant changes in the cranial–caudal position of the heart could be observed. The centroid of the cardiac blood pool showed an excursion of 4.5–16.5 mm (mean 8.5±4.8 mm) in the cranial–caudal direction, with more limited excursion of 1.1–7.0 mm (mean 2.5±2.2 mm) in the horizontal direction and 1.3–3.7 mm (mean 2.4±0.9 mm) in the vertical direction. Conclusion These preliminary data show that the extent of motion involved in respiration is comparable to myocardial wall thickness, and respiratory gating may be considered in order to reduce this effect in the reconstructed images.Keywords
This publication has 19 references indexed in Scilit:
- Rigid-body transformation of list-mode projection data for respiratory motion correction in cardiac PETPhysics in Medicine & Biology, 2005
- Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancerMedical Physics, 2002
- Real-time system for respiratory-cardiac gating in positron tomographyIEEE Transactions on Nuclear Science, 1998
- Non-rigid summing of gated PET via optical flowIEEE Transactions on Nuclear Science, 1997
- Local threshold for segmented attenuation correction of PET imaging of the thoraxIEEE Transactions on Nuclear Science, 1994
- Respiratory kinematics of the upper abdominal organs: A quantitative studyMagnetic Resonance in Medicine, 1992
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989
- Effect of Respiratory Motion on Pulmonary Activity Determinations by Positron Tomography in DogsInvestigative Radiology, 1985
- Influence of Cardiac and Respiratory Motion on Tomographic Reconstructions of the HeartJournal of Computer Assisted Tomography, 1982
- Electrocardiographic Gating in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1979