High-Throughput Mass Spectrometer Using Atmospheric Pressure Ionization and a Cylindrical Ion Trap Array
- 3 December 2004
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 77 (2) , 459-470
- https://doi.org/10.1021/ac048983w
Abstract
The analytical performance of an atmospheric pressure sampling, multiple-channel, high-throughput mass spectrometer was investigated using samples of a variety of types. The instrument, based on an array of cylindrical ion traps, was built with four independent channels and here is operated using two fully multiplexed channels (sources, ion optics, ion traps, detectors) capable of analyzing different samples simultaneously. Both channels of the instrument were incorporated within the same vacuum system and operated using a common set of control electronics. A multichannel electrospray ionization source was assembled and used to introduce samples including solutions of organic compounds, peptides, and proteins simultaneously into the instrument in a high-throughput fashion. Cross-talk between the channels of the instrument occurred in the detection system and could be minimized to 1-2% using shielding between detector channels. In this initial implementation of the instrumentation, an upper mass/charge limit of approximately 1300 Th was observed (+13 charge state of myoglobin) and unit mass/charge resolution was achieved to approximately 800 Th. The rather limited dynamic range (2-3 orders of magnitude for low-concentration analytes) is due to cross-talk contributions from more concentrated species introduced into a different channel. Analysis of mixtures of alkylamines and peptides is demonstrated, but analysis of mixtures with a wide spread in mass/charge ratios was not possible due to mass discrimination in the ion optics. Further refinement of the vacuum system and ion optics will allow the addition of more channels of parallel mass analysis and facilitate applications in fields such as proteomics and metabolomics.Keywords
This publication has 26 references indexed in Scilit:
- Rectilinear Ion Trap: Concepts, Calculations, and Analytical Performance of a New Mass AnalyzerAnalytical Chemistry, 2004
- Multiple sprayer system for high‐throughput electrospray ionization mass spectrometryRapid Communications in Mass Spectrometry, 2002
- Double resonance ejection in a micro ion trap mass spectrometerRapid Communications in Mass Spectrometry, 2002
- Mass spectrometry in combinatorial chemistryMass Spectrometry Reviews, 2000
- Characterization of a serial array of miniature cylindrical ion trap mass analyzersRapid Communications in Mass Spectrometry, 1999
- Evaluated Gas Phase Basicities and Proton Affinities of Molecules: An UpdateJournal of Physical and Chemical Reference Data, 1998
- Miniature, high-resolution, quadrupole mass-spectrometer arrayReview of Scientific Instruments, 1997
- High-pressure effects in miniature arrays of quadrupole analyzers for residual gas analysis from 10−9 to 10−2 TorrJournal of Vacuum Science & Technology A, 1996
- Laser produced ions stored in a cylindrical ion trap and detected in a reflectron time-of-flight mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1994
- Recent improvements in and analytical applications of advanced ion trap technologyInternational Journal of Mass Spectrometry and Ion Processes, 1984