Assessment of five portable blood glucose meters for use in cats
- 1 December 2000
- journal article
- research article
- Published by American Veterinary Medical Association (AVMA) in American Journal of Veterinary Research
- Vol. 61 (12) , 1587-1592
- https://doi.org/10.2460/ajvr.2000.61.1587
Abstract
Objective—To evaluate the clinical and analytic accuracy of 5 portable blood glucose meters (PBGM) in cats, with emphasis on the detection of potential sources of error. Animals—200 cats. Procedure—Venous blood glucose readings from 5 PBGM were compared with the results of a hexokinase reference method. Agreement among methods was determined by error grid analysis and statistical methods. Results—A total of 2,975 PBGM readings and 513 reference values were analyzed. The accuracy of the PBGM varied in different glycemic ranges. The largest differences between PBGM readings and reference values were in the high glycemic range; 4 PBGM underestimated and 1 PBGM overestimated the reference values in most instances. In the low and reference glycemic ranges, the absolute differences between PBGM readings and reference values were small. Despite the analytic differences in accuracy, 4 PBGM had 100% and 1 PBGM had 98.7% of readings in the clinically acceptable values of the error grid analysis. Within- and between-day precisions were good for all PBGM. Significant differences were not detected between readings of EDTA and lithium-heparinized blood and fresh blood without anticoagulant. Compared with these blood types, 1 PBGM had significantly different readings with fluoride anticoagulated blood. In blood samples with a low Hct, all PBGM overestimated glucose concentrations. Sample volumes < 3 µl resulted in inaccurate measurements in 3 PBGM. Conclusions and Clinical Relevance—Performance varied among the 5 PBGM analyzed; however, all PBGM were deemed acceptable for clinical use in cats. (Am J Vet Res 2000;61:1587–1592)Keywords
This publication has 22 references indexed in Scilit:
- Validation of Home Blood Glucose Meters With Respect to Clinical and Analytical ApproachesDiabetes Care, 1998
- Evaluation of Glucocard Memory 2 and Accutrend® Sensor Blood Glucose Meterscclm, 1998
- Accuracy of Devices Used for Self-Monitoring of Blood GlucoseAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1998
- Accuracy, precision and user-acceptability of self blood glucose monitoring machinesDiabetes Research and Clinical Practice, 1997
- Evaluation of a simplified intravenous glucose tolerance test and a reflectance glucose meter for use in catsVeterinary Record, 1997
- A comparison of statistical techniques to evaluate the performance of the glucometer elite® blood glucose meterClinical Biochemistry, 1996
- Accuracy of home blood glucose meters during hypoglycemiaDiabetes Care, 1996
- Long-term Monitoring of the Diabetic Dog and CatVeterinary Clinics of North America: Small Animal Practice, 1995
- Pentamidine-Induced Derangements of Glucose Homeostasis: Determinant roles of renal failure and drug accumulation: A study of 128 patientsDiabetes Care, 1995
- Self-Monitoring of Blood GlucoseDiabetes Care, 1994