Evaluation of the automicrobic system for susceptibility testing of Pseudomonas aeruginosa to gentamicin, tobramycin, and amikacin
- 1 April 1984
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 19 (4) , 502-505
- https://doi.org/10.1128/jcm.19.4.502-505.1984
Abstract
The AutoMicrobic system (AMS; Vitek Systems, Inc., Hazelwood, Mo.) was studied for its ability to produce accurate and precise MIC interpretations for Pseudomonas aeruginosa susceptibility to gentamicin, tobramycin, and amikacin. MICs were determined in parallel on 200 selected P. aeruginosa isolates by using the AMS discrete-integer MIC program AMS p12.ROB for interpretation of the AMS Gram-Negative General Susceptibility Urinary Card, and a reference small-integer broth microdilution test. Parallel AMS and broth microdilution MICs were also replicated for three selected strains of P. aeruginosa for which MICs were representative of the dilution test ranges. For the 200 P. aeruginosa isolates, mean AMS MICs were significantly larger than the reference test mean MICs, coefficients of variation were approximately double those of the reference test, and correlation coefficients were unacceptably low for each antimicrobial agent. MIC replication studies for the three selected P. aeruginosa strains showed comparable AMS and reference mean MICs in the lower portions of the dilution ranges, significantly higher AMS mean MICs in the upper portions, and mean coefficients of variation of 63 and 9.6%, respectively, for replicated AMS and reference MICs. These results indicate that the AMS, in its present stage of development, does not produce acceptable MIC measurements for P. aeruginosa susceptibility to gentamicin, tobramycin, and amikacin.This publication has 22 references indexed in Scilit:
- Evaluation of the automicrobic system gram-positive identification card for species identification of coagulase-negative staphylococciJournal of Clinical Microbiology, 1983
- Comparison of Automated and Rapid Manual Methods for the Same-day Identification of EnterobacteriaceaeAmerican Journal of Clinical Pathology, 1983
- Clinical evaluation of new AMS GSC Plus CardDiagnostic Microbiology and Infectious Disease, 1983
- Use of the automicrobic and enteric-tek systems for identification of EnterobacteriaceaeJournal of Clinical Microbiology, 1982
- Evaluation of the AutoMicrobic system for identification of glucose-nonfermenting gram-negative rodsJournal of Clinical Microbiology, 1982
- A Comparison of Minimum Inhibitory Concentration Values Determined by Three Antimicrobic Dilution Methods for Pseudomonas aeruginosaAmerican Journal of Clinical Pathology, 1981
- Cation Components of Mueller-Hinton Agar Affecting Testing of Pseudomonas aeruginosa Susceptibility to GentamicinAntimicrobial Agents and Chemotherapy, 1980
- The Automicrobic System for urinesJournal of Clinical Microbiology, 1979
- Antibiotic Susceptibility Testing of Pseudomonas aeruginosa: Selection of a Control Strain and Criteria for Magnesium and Calcium Content in MediaThe Journal of Infectious Diseases, 1974
- Effect of the Concentrations of Magnesium and Calcium on the In-Vitro Susceptibility of Pseudomonas aeruginosa to GentamicinThe Journal of Infectious Diseases, 1971