Measurement of trabecular bone mineral density in the distal radius by two gamma-ray computed tomography scanners
- 1 August 1993
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 14 (3) , 269-276
- https://doi.org/10.1088/0967-3334/14/3/005
Abstract
The precision and measurement agreement of two gamma-ray computed tomography forearm bone scanners have been evaluated. The inherent precision in vitro for the Stratec SCT 900 scanner was 0.28% compared with 0.04% for the OSCAR scanner. The measurements were linear over the range of densities found in humans. Increasing thickness of the cortical shell dramatically increased trabecular bone attenuation coefficient (TBAC) for the SCT 900, but only slightly affected results with the OSCAR. Short-term precision of the SCT 900 in 22 women (1.26%) was not as good as that previously reported for the OSCAR (0.50%). The measured TBAC for the same patients was significantly greater with the SCT 900. Corrections for the error introduced by the cortical shell of the radius with the SCT were calculated. Using this correction similar TBAC values could be obtained with both scanners, but the variance of the differences was too great for the results to be used interchangeably. Published bone density data for normal subjects indicate that Scottish women lose more trabecular bone in the two decades after 50 years of age than those in the Federal Republic of Germany.Keywords
This publication has 9 references indexed in Scilit:
- Periphere quantitative Computertomographie zur Knochenmineralmessung mit einem neuen speziellen QCT-ScannerRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1991
- Quantitative γ-ray computed tomography of the radius in normal subjects and osteoporotic patientsThe British Journal of Radiology, 1990
- Differential effect of long-term oestrogen therapy on trabecular and cortical boneMaturitas, 1989
- Correction of scatter in computed tomography images of boneMedical Physics, 1987
- Precision of measurement of bone density with a special purpose computed tomography scannerThe British Journal of Radiology, 1986
- Normal values for trabecular and cortical bone in the radius measured by computed tomographyClinical Physics and Physiological Measurement, 1985
- A gamma-ray computed tomography scanner for the quantitative measurement of bone densityJournal of Biomedical Engineering, 1984
- Quantification of Trabecular Bone with Low Dose Computed TomographyJournal of Computer Assisted Tomography, 1981
- The exponential edge-gradient effect in X-ray computed tomographyPhysics in Medicine & Biology, 1981