Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams
- 2 July 2003
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (14) , 2081-2099
- https://doi.org/10.1088/0031-9155/48/14/304
Abstract
Absolute dosimetry with ionization chambers of the narrow photon fields used in stereotactic techniques and IMRT beamlets is constrained by lack of electron equilibrium in the radiation field. It is questionable that stopping-power ratio in dosimetry protocols, obtained for broad photon beams and quasi-electron equilibrium conditions, can be used in the dosimetry of narrow fields while keeping the uncertainty at the same level as for the broad beams used in accelerator calibrations. Monte Carlo simulations have been performed for two 6 MV clinical accelerators (Elekta SL-18 and Siemens Mevatron Primus), equipped with radiosurgery applicators and MLC. Narrow circular and Z-shaped on-axis and off-axis fields, as well as broad IMRT configured beams, have been simulated together with reference 10 x 10 cm2 beams. Phase-space data have been used to generate 3D dose distributions which have been compared satisfactorily with experimental profiles (ion chamber, diodes and film). Photon and electron spectra at various depths in water have been calculated, followed by Spencer-Attix (delta = 10 keV) stopping-power ratio calculations which have been compared to those used in the IAEA TRS-398 code of practice. For water/air and PMMA/air stopping-power ratios, agreements within 0.1% have been obtained for the 10 x 10 cm2 fields. For radiosurgery applicators and narrow MLC beams, the calculated s(w,air) values agree with the reference within +/-0.3%, well within the estimated standard uncertainty of the reference stopping-power ratios (0.5%). Ionization chamber dosimetry of narrow beams at the photon qualities used in this work (6 MV) can therefore be based on stopping-power ratios data in dosimetry protocols. For a modulated 6 MV broad beam used in clinical IMRT, s(w,air) agrees within 0.1% with the value for 10 x 10 cm2, confirming that at low energies IMRT absolute dosimetry can also be based on data for open reference fields. At higher energies (24 MV) the difference in s(w,air) was up to 1.1%, indicating that the use of protocol data for narrow beams in such cases is less accurate than at low energies, and detailed calculations of the dosimetry parameters involved should be performed if similar accuracy to that of 6 MV is sought.Keywords
This publication has 25 references indexed in Scilit:
- A Monte Carlo based phase space model for quality assurance of intensity modulated radiotherapy incorporating leaf specific characteristicsMedical Physics, 2002
- A Monte Carlo study of radiation transport through multileaf collimatorsMedical Physics, 2001
- Intensity-modulated radiotherapy: current status and issues of interestInternational Journal of Radiation Oncology*Biology*Physics, 2001
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- On the beam quality specification of high‐energy photons for radiotherapy dosimetryMedical Physics, 2000
- Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beamsMedical Physics, 1999
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- A comparison of dosimetry techniques in stereotactic radiosurgeryPhysics in Medicine & Biology, 1996
- Monte Carlo and convolution dosimetry for stereotactic radiosurgeryInternational Journal of Radiation Oncology*Biology*Physics, 1990
- Stopping power data for high-energy photon beamsPhysics in Medicine & Biology, 1986