Quantification of Contrast in Clinical MR Brain Imaging at High Magnetic Field
- 1 July 1985
- journal article
- research article
- Published by Wolters Kluwer Health in Investigative Radiology
- Vol. 20 (4) , 360-369
- https://doi.org/10.1097/00004424-198507000-00005
Abstract
The relative contrast between 2 tissues in a magnetic resonance (MR) image is shown to be quantifiable for any combination of pulse timing parameters, provided the intrinsic parameters are known. Based on multiple inversion-recovery and spin echo images, a region-of-interest T1, T2 and density analysis was conducted at 1.4 T [tesla] in selected patients with diagnosed neuropathology for various brain tissues. The resulting tissue parameters subsequently served to calculate the contrast-to-noise (C/N) ratio for typical tissue interfaces as a function of the operator-variable pulse timing parameters and the data were compared with the images. Although such calculations may be useful as a protocol selection aid, it is obvious that an optimized pulse protocol can only be established for a single tissue interface. A T2-discriminating pluse sequence like Carr-Purcell-Meiboom-Gill with long repetition time, generally advocated as clinically most effective, may not always be ideal.Keywords