Finite Element Analysis of a Total Knee Replacement by Using Gauss Point Contact Constraints
- 1 March 1996
- journal article
- review article
- Published by SAGE Publications in Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
- Vol. 210 (1) , 51-63
- https://doi.org/10.1243/pime_proc_1996_210_390_02
Abstract
Finite element methods have been applied extensively and with much success in the analysis of orthopaedic hip and knee implants. Very recently a burgeoning interest has developed, in the finite element community, in how numerical models can be constructed for the solution of problems in contact mechanics. New developments in this area are of paramount importance in the design of implants for orthopaedic surgery. Modern techniques are described for finite element contact analysis and applied to two problems of stress analysis in a plastic tibial component. In the former, results are compared with a previous finite element analysis and with Hertzian solutions. In the latter, an estimate of the extent of convergence of the finite element solutions is provided.Keywords
This publication has 20 references indexed in Scilit:
- Pre-Clinical Testing of Hip Prosthetic Designs: A Comparison of Finite Element Calculations and Laboratory TestsProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1993
- Parametric Modelling of a Knee Joint ProsthesisProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1993
- A general finite element approach for contact stress analysisInternational Journal for Numerical Methods in Engineering, 1992
- An augmented lagrangian treatment of contact problems involving frictionComputers & Structures, 1992
- Polyethylene wear in uncemented knee replacementsThe Journal of Bone and Joint Surgery. British volume, 1992
- Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prosthesesThe Journal of Bone and Joint Surgery. British volume, 1992
- A Numerical Model of the Load Transmission in the Tibio-Femoral Contact AreaProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1990
- Polyethylene wearJournal of Bone and Joint Surgery, 1988
- A radiographic study of bearing movement in unicompartmental Oxford knee replacementsThe Journal of Bone and Joint Surgery. British volume, 1987
- The Deformation and Wear of Plastic Components in Artificial Knee Joints — An Experimental StudyEngineering in Medicine, 1981