NEMA NU 2-2007 performance measurements of the Siemens Inveon™ preclinical small animal PET system
Open Access
- 25 March 2009
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (8) , 2359-2376
- https://doi.org/10.1088/0031-9155/54/8/007
Abstract
National Electrical Manufacturers Association (NEMA) NU 2-2007 performance measurements were conducted on the Inveon preclinical small animal PET system developed by Siemens Medical Solutions. The scanner uses 1.51 x 1.51 x 10 mm LSO crystals grouped in 20 x 20 blocks; a tapered light guide couples the LSO crystals of a block to a position-sensitive photomultiplier tube. There are 80 rings with 320 crystals per ring and the ring diameter is 161 mm. The transaxial and axial fields of view (FOVs) are 100 and 127 mm, respectively. The scanner can be docked to a CT scanner; the performance characteristics of the CT component are not included herein. Performance measurements of spatial resolution, sensitivity, scatter fraction and count rate performance were obtained for different energy windows and coincidence timing window widths. For brevity, the results described here are for an energy window of 350-650 keV and a coincidence timing window of 3.43 ns. The spatial resolution at the center of the transaxial and axial FOVs was 1.56, 1.62 and 2.12 mm in the tangential, radial and axial directions, respectively, and the system sensitivity was 36.2 cps kBq(-1) for a line source (7.2% for a point source). For mouse- and rat-sized phantoms, the scatter fraction was 5.7% and 14.6%, respectively. The peak noise equivalent count rate with a noisy randoms estimate was 1475 kcps at 130 MBq for the mouse-sized phantom and 583 kcps at 74 MBq for the rat-sized phantom. The performance measurements indicate that the Inveon PET scanner is a high-resolution tomograph with excellent sensitivity that is capable of imaging at a high count rate.Keywords
This publication has 21 references indexed in Scilit:
- Imaging of Weak-Source Distributions in LSO-Based Small-Animal PET ScannersJournal of Nuclear Medicine, 2007
- Performance Measurement of the microPET Focus 120 ScannerJournal of Nuclear Medicine, 2007
- Performance evaluation of the Philips MOSAIC small animal PET scannerEuropean Journal of Nuclear Medicine and Molecular Imaging, 2006
- Radiation dose estimate in small animal SPECT and PETMedical Physics, 2004
- Performance evaluation of the microPET R4 PET scanner for rodentsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- A 3D HIDAC-PET camera with sub-millimetre resolution for imaging small animalsIEEE Transactions on Nuclear Science, 1999
- MicroPET: a high resolution PET scanner for imaging small animalsIEEE Transactions on Nuclear Science, 1997
- Exact and approximate rebinning algorithms for 3-D PET dataIEEE Transactions on Medical Imaging, 1997
- Design and engineering aspects of a high resolution positron tomograph for small animal imagingIEEE Transactions on Nuclear Science, 1994
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990